Time Interval (Single And Double Line); Time Interval Average; Time Interval Average Measurement - Racal Instruments 9901 Maintenance Manual

Universal counter-timer
Table of Contents

Advertisement

TIME
INTERVAL
AVERAGE
3.11
The
reference
frequency
is
10
MHz,
thus
the
shortest
available clock
unit
is
100
ns,
which
places
a
limit
on
the resolution
available
for
such
measurements
such
as
narrow
pulse widths.
Greater
resolution
is
achieved by
the 'averaging'
technique,
which
functions
as
follows:-
3.12
Referring
to
Figure
3.4
the
clock
input
to
the counter has
to
pass
through
two
gates
in series
(gates 'A'
and
'B'
in
the
diagram).
The
successive
Start
and
Stop
pulses
open and
close
Gate
'A'
allowing 10
MHz
clock
pulses
to
pass to
Gate
'B'
during
each
'gate
open' period.
Gate
'B'
is
held
open by
the
time base
for
the selected time
(10*^)
and
the
clock
pulses are totalled
in
the
counter.
At
the
end
of the selected time
base period
Gate
'B'
closes
and
the
count
is
displayed
as
the
average
time
interval
.
NOTE:
To
obtain
the
improved
resolution
from time
interval
averaging,
the
recurrence
frequency
of the timing input signal
must
not
be harmonically
related
to
the
frequency
standard of the counter.
Time
Interval
Average
Measurement
Fig.
3.
4
9901 Vol.
2
3-5