ABL800 FLEX Reference Manual
Gas mixtures (Gas 1 and Gas 2)
For calibration of the pCO
Use
Cylinder types
The following types of Gas 1 cylinders are used depending on the geographical
location of the analyser:
Cylinder Volume
WARNING/
Pressurized container. Non-flammable compressed gas. Do not breathe gas. Gas
CAUTION:
mixtures containing less than 19.5 % oxygen may cause suffocation. Protect from
sunlight and do not expose to temperatures exceeding 50 C (122 F). Store and
use with adequate ventilation. Keep away from oil and grease. Do not refill.
NOTE:
The exact composition of each gas mixture is given in the barcode on the gas
cylinder label. The barcode is entered by the barcode reader or manually.
Stability
Gas 1 and Gas 2 are stable for 25 months from the date of filling.
Storage
The gas cylinders should be stored between 2 - 32
-
2
EU
1 L
Gas Volume
10 L
Fill Pressure
140 psi
o
at 25
C
(10 bar)
Composition
and pO
electrodes.
2
Gas 1
USA
Japan
1 L
1 L
33 L
25 L
500 psi
375 psi
(34 bar)
(26 bar)
19.76 % O
, 5.60 % CO
2
74.64 % N
2
7. Solutions and gas mixtures
Gas 2
1 L
10 L
140 psi
(10 bar)
< 0.04 % O
, 11.22 % CO
2
2
88.78 % N
o
o
o
o
C (36 - 90
F).
2
2
7-7
Need help?
Do you have a question about the ABL800 FLEX and is the answer not in the manual?