prietest TOUCH
3.3. Technical specification of prietest TOUCH
Human Machine Interface
Linear measurement range
Photometric Accuracy
Drift
Photometric Linearity
Optical measurement
Filters
Type of filter
Wave Length
Half Bandwidth
Filter Selection
Flow Cell
Sipping Volume
Flow cell Volume
Sipping Mode
Cuvette Volume
Dry Block Incubator
Number of cuvettes
Temperature
Temperature of cuvette/flow cell Block
Method
Temperature
Light source
Warm up Time
Display
Printer
Memory
Storage Capacity
Analysis Mode
Concentration Calculation
RS232 serial port
Power
Wattage
Voltage
Operating Position
Operating Conditions
Temperature
Relative Humidity
Storage Condition
Temperature
Relative Humidity
Enclosure
Size (cm)
Weight
Version No.: 2.622A
Effective Date: July, 2010
TOUCH PANEL / KEYPAD
0.000 to 3.000 Absorbance Units (A).
± 2% or 0.007 whichever is higher, from 0 to 1.5 A
± 3% from 1.5 A to 3.0 A
<0.007 A/hr
2.2 A
Photodiode
Interference
340, 405, 510, 546, 578 and 630 nm & Two Optional
10nm ± 2nm
Automatic by Stepper Motor.
300 to 1000 µl
18µl
Automatic by specially designed Peristaltic Pump
500 µl
15 Test Tubes
0
37
C
By Peltier effect
0
25, 30 and 37
C
Tungsten Halogen
90 Sec
5" Graphic LCD, Negative Blue, STN
Built in thermal printer
64 KB – Non volatile RAM with Battery back up
250 Open tests, with 30 QC results / test
(Normal and abnormal controls) and more than
2500 Patient results with patient ID (6 digits)
Absorbance
End Point
Ratio
Coagulation
By factor or by Standard
9600 baud, 1 start, 8 data, 1 stop, no parity bits
50 watts
115 – 230 Volts ± 10 %, 60 – 50 Hz
On horizontal, flat, rigid and vibration free s urface
0
From + 18
C to + 35
Up to 85 %
0
From –10
C to + 60
Up to 85 %
ABS Fire retardant
30 X 38 X 13.5 (l X b X h)
5.5 Kgs. (Approx)
Page 8 of 52
USER MANUAL
Kinetic
Differential
Fixed Time
0
C
0
C