Do you have a question about the prietest TOUCH and is the answer not in the manual?
Questions and answers
Satendra
May 7, 2025
How to change forget password
1 comments:
Mr. Anderson
May 18, 2025
To reset the password for ROBONIK prietest TOUCH:
1. Go to the PASSWORD UTILITY screen. 2. Select option 1: Enable / Disable. 3. Enter the required user name (e.g., "DAVID") and touch ENTR. 4. To set a new password, select option 2: Set Password. 5. Enter the new password and confirm.
Note: A message “password enabled” will appear at the bottom of the screen when successful.
This answer is automatically generated
Christian Kagheni
June 2, 2025
I forgot the password of my device what can I do
Christian Kagheni
June 2, 2025
j'ai oublie le mot de passe de mon appareil qu'est ce que je dois faire
Satendra
May 7, 2025
Who is the seller of robonik analyser in nepal
rajesh
March 30, 2025
Is the meril and erba company reagents use to robonic analyser for tests
USER MANUAL prietest TOUCH TABLE OF CONTENTS 1. GENERAL INFORMATION ......................3 1.1. W : ....................3 ARRANTY NFORMATION 1.2. T : ......................3 ECHNICAL ERVICE 1.3. D :....................3 ISPOSAL NSTRUCTION 1.4. C : ........................3 ONTACTS 2. GENERAL SAFETY WARNINGS ..................... 3 2.1.
Page 3
USER MANUAL prietest TOUCH 8.3. UTILITIES SCREEN....................... 3 8.3.1. To ENTER/CHANGE Date and Time....................... 3 8.3.2. Password Utility..............................3 8.3.3 WASH F.C ................................3 8.3.4 Mode:..................................3 8.3.5 Pump Cal:................................3 8.3.6 Program Filter ..............................3 8.3.7. To disable or enable the Printer........................3 8.3.8 To Enter Clinic / Doctor Name:.........................
ROBONIK will at its discretion repair or replace parts, which may be found defective in the warranty period. The warranty does not include any responsibility for direct or indirect personal and material damages, caused by improper use or maintenance of the instrument.
USER MANUAL prietest TOUCH 2. GENERAL SAFETY WARNINGS 2.1. Danger – warnings symbols: The following symbols are used to inform the user of the safety rules. This symbol indicates generic danger. It means that, serious damage can occur to the operator if described precautions are not observed.
USER MANUAL prietest TOUCH 2.2. Use of the instrument: Ø The instrument has to be used for the designed purposes under specified conditions, following proper procedures and safety rules, by qualified personnel. Ø THIS MANUAL CONTAINS INSTRUCTIONS FOR OPERATION BY QUALIFIED PERSONNEL Ø...
3. INTRODUCTION 3.1 DESCRIPTION: prietest TOUCH is a programmable Biochemistry Analyzer with a user-friendly touch screen. It measures the optical densities of samples and it uses algorithm to calculate results, which are used for biochemical investigation. It is a photometer operating in the visible range.
USER MANUAL prietest TOUCH 3.3. Technical specification of prietest TOUCH Human Machine Interface TOUCH PANEL / KEYPAD Linear measurement range 0.000 to 3.000 Absorbance Units (A). Photometric Accuracy ± 2% or 0.007 whichever is higher, from 0 to 1.5 A ±...
USER MANUAL prietest TOUCH 4. PACKING, TRANSPORT, AND STORAGE 4.1. GENERAL WARNINGS: Instrument has to be decontaminated before packing for transportation. 4.2. PACKING: Packaging is needed whenever the instrument is to be transported or shipped by courier or other means.
TOUCH 5. INSTRUMENT DESCRIPTION 5.1. Touch Sensitive screen prietest TOUCH provides a Touch sensitive LCD panel and a KEYPAD for easy user interface. The Menus are displayed; the text of the parameter forms the TOUCH ZONE. Touch screen Layout For Example: In Kinetic Mode.
USER MANUAL prietest TOUCH 5.2. Perspective View Front View: Incubator Printer Flowcell Aspiration Switch Touch Screen Keypad Rear View: Internal Printer Port (Optional) Contrast Pot ON / OFF Switch Waste Outlet Peristaltic Pump External Printer Port Cooling Fan RS 232...
Page 12
USER MANUAL prietest TOUCH 5.3. KEYPAD 36.50 MAIN MENU 1. Test Operations 2. Technical Diagnosis 3. Utilities 4. Quality control 5. View Stored Results 6. Communication The FEED key is used to advance the printer paper by 1 line. PRINT key is used to take the print.
USER MANUAL prietest TOUCH 5.4. Probe The instrument is provided with a probe to aspirate the sample into the flow cell. The push button (Aspiration Switch) has to be pressed to activate the peristaltic pump. 5.5. Peristaltic pump The instrument is provided with a peristaltic suction pump. The aspiration push button switch activates the peristaltic pump.
USER MANUAL prietest TOUCH PRINCIPLE OF WORKING: Below is the diagram representing the main functional elements of the instrument. White light produced by the lamp is focused into a beam by lens and passes through the interference filter and monochromatic light beam falls on the sample. Part of the light is absorbed by the sample, the remaining is transmitted.
Page 15
TOUCH 5.7. Printer A) Internal Printer (Thermal Printer) prietest TOUCH comes with a built in 20 column Thermal Printer. User has to take proper care to handle this delicate instrument. TIPS FOR CAR EFUL USAGE OF PRINTER Do not pull the paper when loaded 1.
PRIETEST TOUCH PRIETEST TOUCH 36.50 36.50 Clinic Name MENU MAIN MENU ROBONIK (INDIA) PVT. LTD. 1. Test Operations Plot No. A-374, TTC, 2. Technical Diagnosis MIDC Industrial Area, Mahape Select Test PRNT 3. Utilities Navi Mumbai– 400 710...
Page 17
USER MANUAL prietest TOUCH 7. PRECAUTIONS λ Keep the place dry and clean. λ Check all the grounding wires properly. Repeat the readings, if Absorbance is more than 2.0 A. Use original Packing for transportation. λ Use clean Cuvettes. Check the blank absorbance of the cuvette at regular intervals.
USER MANUAL prietest TOUCH 8. MAIN MENU 8.1. TEST OPERATIONS A) Programming/ADDING a NEW TEST When the unit is first turned on, Test List screen / Main Menu Screen appears on the display. Main Menu screen 36.45 33636636. 36.50 MAIN MENU 1.
Page 19
USER MANUAL prietest TOUCH B) Recalling / Running a Saved TEST (Operating the programmed Tests) The Programmed/Saved Test can be recalled/selected; either by Name, Mode, Number or List Tests when the unit is first turned on, Test List screen / Main Menu Screen appears on the display.
USER MANUAL prietest TOUCH C) List Tests: • Displays max. of 10 Tests/ Page . A “List tests” screen would look like the screen shown below, First column : “Test Number”, second column : “Test Name ” ,Third column :Number of times the test has been run, and fourth column : “...
USER MANUAL prietest TOUCH 8.2.4. Temperature Setting To verify the set temperature. 36.74 36.71 TECHNICAL DIAGNOSIS 1. Clear Memory MAIN MENU 2. Home Filter Wheel 1. Test Operations 3. Select Filter 2. Technical Diagnosis 4. Temperature Setting 3. Utilities 5. Lamp Amplitude 4.
USER MANUAL prietest TOUCH 8.2.6. Printer Self Test Routine test to check printer. When selected, 36.40 36.40 TECHNICAL DIAGNOSIS MAIN MENU 1. Clear Memory 1. Test Operations 2. Home Filter Wheel 2. Technical Diagnosis 3. Select Filter 3. Utilities Temperature Setting 4.
USER MANUAL prietest TOUCH 8.3.3 WASH F.C Wash Flow cell is used to wash the flow cell .Selecting “Wash F.C” will allow the aspiration of water or cleaning solution in the Flowcell. 36.40 UTILITIES Ver (Version No) 36.68 MAIN MENU 1.
USER MANUAL prietest TOUCH 8.3.5 Pump Cal: The option of “Pump Cal” allows the operator to adjust the volume being aspirated through the Flowcell. The adjustment of the peristaltic pump has two functions: To adjust the volume of liquid aspirated into the flow cell. The volume of liquid must be enough to rinse Flowcell sufficiently, but not more than the reaction mixture;...
USER MANUAL prietest TOUCH 8.3.6 Program Filter The option of “Prog. Fil.” allows the operator to enter the wavelength of the filters in 7 and 8 position. This option is to be used only in case of instruments where it is required to filters in 7 and 8 position.
USER MANUAL prietest TOUCH 8.4. QUALITY CONTROL Quality Control is a process that checks an instrument or testing site to make sure it is reporting accurate results on patients. The reproducibility of a result from a testing site or instrument should fall within a certain range. Control solutions of known values are often times used for checking quality control.
USER MANUAL prietest TOUCH 8.6. Communication: The instrument is equipped with an RS232 serial port for PC configuration (user-computer interface.). A cable is available to link the instrument to PC Communication will only start when both ends detect the presence of an active terminal or device.
USER MANUAL prietest TOUCH 9. PROGRAMMING MODES 9.1. ABSORBANCE The instrument measures the blank transmittance and the sample transmittance, subtracts them and calculates the sample absorbance ( m onochromatic mode).It is possible to repeat the measure with a different wavelength(Filter 2/ secondary filter) and consider the difference between the two measures.
USER MANUAL prietest TOUCH 9.2. FIXED TIME Change in absorbance of sample is taken at programmed time interval and concentration is calculated either from the factor fed by the user or using standard. PROGRAMMING / ADDING a NEW TEST (Refer chapter 8)
USER MANUAL prietest TOUCH 9.3. KINETIC Multiple readings are taken at set temperature, at regular intervals and change in absorbance per minute is calculated. Concentration is calculated from the factor fed by the user or by using standard. PROGRAMMING / ADDING a NEW TEST (Refer chapter 8)
USER MANUAL prietest TOUCH 9.4. END POINT The instrument reads absorbance of the sample and calculates concentration using fed Factor or calculates the factor from concentration of the standard. PROGRAMMING / ADDING a NEW TEST (Refer chapter 8) Name: , Mod: END-P 36.45...
USER MANUAL prietest TOUCH 9.5. DIFFERENTIAL Differential of Sample and Sample Blank is taken. Concentration is calculated either from the factor fed by using standard. PROGRAMMING / ADDING a NEW TEST (Refer chapter 8) Name: , Mod: DIFF 36.42 Pri. : 340 , Sec.
USER MANUAL prietest TOUCH 9.6. RATIO PROGRAMMING / ADDING a NEW TEST (Refer chapter 8) Name: , Mod: RATIO 36.54 Pri. : 340 , Sec. : 0 Temp: 37C , KF: 1.000 Vol: 300ul , Unit: No –Unt Lag : 0...
USER MANUAL prietest TOUCH 9.7. COAGULATION This mode is used to report the Prothrombin Time (PT) results based on the ISI (International Sensitive Index) of the thromboplastin reagents and INR (International Normalised Ratio). The INR is (ISI) calculated using formula INR = R , where ISI = Lot specified ISI value of reagent and R= Patient PT/ Normal PT.
USER MANUAL prietest TOUCH 10. GENERAL FUNCTIONS 10.1. Enter Test Name Touch” Name” under the selected mode on the Test Screen. For example :In ENDPOINT mode Name: , Mod: END-P 36.50 Name : Mod:END-P Pri. : 340 , Sec. : 0 Temp: 37C , KF: 1.000...
USER MANUAL prietest TOUCH 10.3. Filter selection. To select the interferential filter required for the test being programmed. For example :In ENDPOINT mode. Name: , Mod: END-P 36.70 Pri. : 340 , Sec. : 0 Temp: 37C , KF: 1.000 Touch “Pri”...
USER MANUAL prietest TOUCH 10.4. Selection of Units For example :In ENDPOINT mode. Name: , Mod: END-P 36.45 Pri. : 340 , Sec. : 0 Touch “UNIT”. Temp: 37C , KF: 1.000 On proper selection the analyser Vol: 300ul , Unit: No –Unt...
USER MANUAL prietest TOUCH 10.8. Lag time Delay interval (in seconds) before readings are executed. For example: In ENDPOINT mode. Name: , Mod: END-P 37.45 Pri. : 340 , Sec. : 0 Touch” Lag”. Temp: 37C , KF: 1.000 On proper selection the analyser Vol: 300ul , Unit: No –Unt...
USER MANUAL prietest TOUCH 10.10. Limit set (End Point, Differential & Ratio) Name: , Mod: END-P 37.25 Rgnt. Linearity: 0.000 Pri. : 340 , Sec. : 0 Temp: 37C , KF: 1.000 Vol: 300ul , Unit: No –Unt 37.12 Lag : 0 , Read: 0 Rgnt.
Page 45
USER MANUAL prietest TOUCH Limit set (Fixed Time) Init O.D : High or low absorbance level. This limit is entered for those samples which may exceed the limits of the reagent system. Max Delta :Upper limit for the Delta Rgnt. Linearity: as per given in a reagent kit.
USER MANUAL prietest TOUCH 10.11. Factor / K-Factor / Standard selection Selection of Standard or Factor prietest TOUCH has the flexibility to calculate the results by 1) Factor Method 2) Single Standard/ Multi-standard For example in kinetic mode, Name: , Mod: KIN 36.72...
Page 47
USER MANUAL prietest TOUCH Name: , Mod: KIN 36.59 Touch “Concen”. Pri. : 340 , Sec. : 0 On proper selection, the analyser Temp: 37C , KF: 1.000 responds with blinking of the parameter Vol: 300ul , Unit: No –Unt...
Page 48
USER MANUAL prietest TOUCH Name: , Mod: KIN 36.54 Pri. : 340 , Sec. : 0 Touch “K-Factor”. Temp: 37C , KF: 1.000 On proper selection the analyser Vol: 300ul , Unit: No –Unt responds with blinking of the parameter...
Page 49
USER MANUAL prietest TOUCH 10.12. QC Normal/Abnormal values QC Normal Values QC Abnormal Values QCNH - Quality control Normal high value QCABH – Quality control Abnormal high value QCNL - Quality control Normal low value QCABL – Quality control Abnormal Low value For example :In ENDPOINT mode.
USER MANUAL prietest TOUCH 12. DECONTAMINATION 12.1. Decontamination Procedure • If the instrument is to be shipped after being exposed to potentially hazardous material, it should be decontaminated. The following procedure outlines the method of decontaminating the instrument before packaging and shipment.
USER MANUAL prietest TOUCH 13. SAFETY CLEARANCE CERTIFICATE Please complete all information requests on this form prior to returning the instrument to the manufacturer or your local distributor for servicing, repairs or return. Thank you for your co- operation. Customer...
Need help?
Do you have a question about the prietest TOUCH and is the answer not in the manual?
Questions and answers
How to change forget password
To reset the password for ROBONIK prietest TOUCH:
1. Go to the PASSWORD UTILITY screen.
2. Select option 1: Enable / Disable.
3. Enter the required user name (e.g., "DAVID") and touch ENTR.
4. To set a new password, select option 2: Set Password.
5. Enter the new password and confirm.
Note: A message “password enabled” will appear at the bottom of the screen when successful.
This answer is automatically generated
I forgot the password of my device what can I do
j'ai oublie le mot de passe de mon appareil qu'est ce que je dois faire
Who is the seller of robonik analyser in nepal
Is the meril and erba company reagents use to robonic analyser for tests