CAIRE HELiOS Marathon H850 Home Use Manual page 49

Portable oxygen system carrier bag
Hide thumbs Also See for HELiOS Marathon H850:
Table of Contents

Advertisement

Available languages

Available languages

  • ENGLISH, page 1
Troubleshooting
7. Frost forms on the oxygen supply line connected
8. The H850 Portable does not disengage easily from
WARNING:
connector when disengaging the Portable. Contact
with liquid oxygen can cause personal injury.
46
to the Reservoir.
Frost on the supply line is most likely the result of a
leaky connection or a split in the line. If the connec-
tion at the Reservoir outlet is leaking, keep tightening
the nut until the leak stops. If the supply line is split or
damaged, disconnect the line from the Reservoir and
contact your home care provider.
the Reservoir after filling.
Do not bend over the Reservoir fill
The H850 Portable and Reservoir fill connectors may
have become frozen. Do not use force. Allow a few
minutes for the frozen parts to warm, then disengage
the H850 Portable when the ice has melted. To pre-
vent the Reservoir and H850 Portable from freezing
together, always wipe the male fill connector on the
Reservoir and the female fill connector on the H850
Portable with a clean, dry, oil-free cloth before filling.
H850 Home Use Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents