Troubleshooting
3. Excessive frost and hissing.
4. The H850 Portable stops pulsing during use.
44
Some hissing and frosting on the H850 Portable is
normal. If, however, the hissing sound from the
H850 Portable continues after a fill or anytime
during usage, combined with excessive frosting
on the H850 Portable, stop using the H850
Portable and contact your Homecare Provider.
Excessive frost formation on the H850 Portable
30 minutes after a fill while the Flow Control
knob is in the OFF position may be a result of
possible system malfunction. Stop using the H850
Portable and contact your Homecare Provider for
service.
Ensure that both connections of the dual-lumen
cannula are firmly attached to the Oxygen Outlet
and Sensor connectors.
Change the cannula if water droplets are present
from humidified exhaled gas. Replacing the
cannula with a dry one enhances the ability of the
H850 Portable to pulse.
Ensure that the cannula is not kinked or pinched.
Ensure that the cannula tips remain in your
nostrils and do not slide to one side.
Ensure that there is oxygen in the H850 Portable.
With the cannula on, close your mouth and
breathe only through your nose to verify that the
H850 Portable has stopped pulsing.
H850 Home Use Guide