Technical Malfunctions - Ivoclar Vivadent Programat P300 Operating Instructions Manual

State-of-the-art furnace for dental applications
Table of Contents

Advertisement

Index Category Error
24
System
Checksum error in the
firing program memory
25
System
Write error in the
firing group memory
26
System
Read error in the
firing group memory
27
System
Checksum error in the
firing group memory
28
System
OT1 <> OT2
plausibility check:
OT1 = OT2 +/- 10°C
29
System
Temperature in the furnace
base is lower than 1 °C
30
System
Program start blocked
31
System
Incorrect time settings
(date / time)
33
Entry
HV > H (H2)
35
Entry
"Share of the holding time
with vacuum" is activated,
but Vx2 does not
correspond to Tx or Tx+1
36
System
Supply voltage outside the
acceptable range
37
System
Start-up aborted due
to an error
38
System
Brief power failure during a 702
firing program in progress
39
System
Power failure during a
firing program from the
memory stick in progress –
memory stick no longer
present.
40
System
Prolonged power failure
during an onvernight
program in progress
41
System
Reading and processing
supply voltage
42
System
Reading the power frequency 706
43
System
Incorrect supply voltage
44
System
Final vacuum value not
reached
45
System
Vacuum drop
46
System
The vacuum does not
increase (self-test)
47
System
Temp. EXTERNAL T-SENSOR 1010
excess temperature
(> 1225°C)
48
System
Write error in the furnace
configuration data memory
49
System
Read error in the furnace
configuration data memory
50
System
Checksum error in the
furnace configuration
data memory
51
System
Write error in the furnace
operational data memory
52
System
Read error in the furnace
operational data memory
53
System
Checksum error in the
furnace operational data
memory
ERR
Conti-
Error Message Text
No.
nuation
possible
45
Invalid checksum of the memory for firing program data - the firing program data are written to the internal
memory using the original values.
46
Error while writing firing group data to the internal memory.
47
Error while reading the firing group data from the internal memory.
48
Invalid checksum of the memory for firing group data - the firing group data are written to the internal
memory using the original values.
54
nein
Error in the temperature measuring circuit! Contact your local Ivoclar Vivadent Service Center.
**,***
56
Temperature in the furnace base is lower than 1 °C. Bring the furnace base to a higher operating temperature.
103
Starting a program is not possible due to a technical malfunction.
107
The clock settings are incorrect. Please set a correct date and a correct time!
110
Enter a lower value for HV or a higher value for H (H2)
120
Activate the vacuum during the holding time Tx or deactivate HV..
700
The supply voltage is outside the acceptable range. Check the supply voltage.
701
The self-test of the furnace was interrupted by an error. It is not possible to work with the furnace! Switch
***
the furnace off and on again, once the error has been rectified.
A firing program in progress was interrupted by a brief power failure. The program is continued!
703
A firing program in progress (started from the USB memory stick) was interrupted by a power failure.
The program could not be continued, since the USB memory stick is no longer present!
704
An overnight program (firing program) in progress was interrupted by a prolonged power failure.
The overnight program is continued!
705
no
Error during measuring the supply voltage. Contact your local Ivoclar Vivadent Service Center.
**,***
Error during measuring the supply voltage. Contact your local Ivoclar Vivadent Service Center.
707
The furnace is operated with the incorrect supply voltage. Make sure that the furnace is operated with the
supply voltage indicated on the rating plate.
800
The required final vacuum value cannot be reached. Check the vacuum pump.
801
An unacceptable vacuum drop has occurred.
802
No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no
contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?
Is the fuse F1 o.k.?
Temperature channel EXTERNAL T-sensor excess temperature
1011
Error while saving furnace configuration data to the internal memory.
1012
Error while reading the furnace configuration data from the internal memory.
1013
Invalid checksum of the memory - furnace configuration data are written into the internal memory using
the original values.
1014
Error while saving the furnace operational data to the internal memory.
1015
Error while reading the furnace operational data from the internal memory.
1016
Invalid checksum of the memory – furnace operational data are written into the internal memory using
the original values.