Gamry Reference 3000 Operator's Manual page 26

Potentiostat/galvanostat/zra
Hide thumbs Also See for Reference 3000:
Table of Contents

Advertisement

Chapter 3 -- Instrument Circuitry--Reference 3000 Schematic/Block Diagrams
Notes for Figure 3-4:
Two ADC Channels are dedicated to the potentiostat's voltage and current signals. This diagram shows
the signals that can be connected to the third channel.
The Aux ADC BNC input is a differential input. Some of this input's characteristics can be changed by
either jumpers or CMOS switches. Early Reference 3000's use jumpers to change the characteristics,
while later units (shipped after the middle of 2009) will use CMOS switches under software control to
change the characteristics. Contact Technical Support at Gamry.com if you are uncertain which type
of unit you have.
Jumper configured Reference 3000's are shipped configured for input impedance of 100 kΩ and
unfiltered operation. The jumpers shown in this diagram allow it to be configured for high input
impedance and/or input filtering.
The thermocouple input allows for connection of a K Type thermocouple. Note that this circuit must
be calibrated to obtain reasonable accuracy. A section of the Reference 3000 calibration procedure
allows for user calibration of this input.
The other two inputs to the Aux channel can be used to measure the AE (Auxiliary Electrometer) signal
or a Control Amp voltage signal.
Figure 3-4
Aux Channel Input Switching
3 - 6

Advertisement

Table of Contents
loading

Related Products for Gamry Reference 3000

Table of Contents