Page 1
Snore Stopper ASG 341 Snore Stopper ASG 341 INSTRUCTION MANUAL • • Snore stopper with electric stimulation current • • 1 channel system with 2 adhesive electrodes on the inside of the device • • Flexible wristband, individually adjustable up to 22 cm wrist size •...
TABLE OF CONTENTS No. Topic Page Definition of symbols Basic information How does a snore stopper work? Information about snoring Safety instructions General safety instructions Usage/environment for which the snore stopper is suited Usage/environment for which the snore stopper is not suited Application for which the snore stopper is suited Application for which the snore stopper is not suited Usage by children and adolescents...
14.0 Terms of warranty Dear purchaser, Congratulations on your purchase of your new snore stopper ASG 341 and thank you for your trust. To ensure optimal functionality and operation of your snore stopper, please first read the instruction manual before using the device. This guarantees a long useful life of...
BASIC INFORMATION 1.0 Definition of symbols The safety symbols shown in this instruction manual contain information relating to the correct use of the snore stopper and your safety. The symbols refer to the following content: Read and follow the instruction manual. Warning/Danger: If not used correctly, serious or fatal injuries and damage may occur! The instructions must be followed at all times!
SAFETY INSTRUCTIONS 3.0 Safety instructions General safety instructions 3.1.1 The snore stopper may not be repaired, used or modified (changed) by the users themselves in case of a defect. If used incorrectly, the stimulation current may cause pain, injuries and burns. 3.1.2 If, during the use of the snore stopper, skin alterations, pain, swellings, discomfort or other anomalies occur, you must stop the therapy immediately and seek...
SAFETY INSTRUCTIONS Usage/environment for which the snore stopper is not suited 3.4.1 The snore stopper may not be used at the same time with other medical and electric devices of any kind. 3.4.2 Do not use or wear the snore stopper while bathing, showering or in any other environment with high air humidity.
SAFETY INSTRUCTIONS 3.6.2 Please consult your doctor before using the snore stopper in the following circumstances: a. acute diseases, b. sleep disturbances, c. infectious diseases, d. fever, e. blood pressure problems, f. skin diseases, g. after an accident, h. nausea or dizziness, i.
If the device is not used for a prolonged period, remove the batteries from the device. 4.2.4 If the snore stopper ASG 341 is intended for trade or commercial use, a technical safety inspection is necessary every 2 years in accordance with § 6 MPBetreibV [German Medical Devices Operator Ordinance]. The technical safety inspection must be carried out by a company specialised in medical devices.
SCOPE OF DELIVERY/DISPOSAL 5.0 Scope of delivery/contents of packaging 1 snore stopper ASG 341 1 instruction manual 3 x adhesive electrodes 3 x alcoholic cleaning 1 x 1.5 V battery AAA wipes 6.0 Information about noise sources, noise scale up to 70 dB...
CHANGING THE BATTERIES 8.0 Changing the batteries and information about batteries Place 1 battery (type AAA) in the device as shown in the figure. Observe the correct polarity (+ and – pole). Type of battery: Only use alkaline batteries for the snore stopper. Rechargeable batteries (batteries: NiMH, NiCd) may not be used.
Page 11
APPLICATION OF THE DEVICE 10.0 Application of the snore stopper Step 1: Before attaching the adhesive electrodes, clean the two black electrode pads with the enclosed alcoholic cleaning wipe (see image 5) or a commercially available alcoholic cleaning wipe. Image 5 Image 5 Image 5 Step 2:...
DESIGNATION/FUNCTION 9.0 Designation and function of the snore stopper Image 3 Image 4 Test button: Press the test button a total of 3 consecutive times to test the set intensity. LED lamp: The green or red LED lamp flashes when the device is switched on. If the red LED flashes, there is no contact between the adhesive electrodes and the skin surface.
Page 13
APPLICATION OF THE DEVICE Step 9: Switch on the device. Turn the on/off switch to intensity level 2 (image 13). If skin contact is good, the green LED flashes once. If skin contact is not good, the red LED flashes once (image 15). In this case, tighten the band a little and repeat step 9 until the green LED lights up.
TECHNICAL PROBLEMS 11.0 Technical problems, troubleshooting Fault Cause Solution The batteries are inserted, There could be foreign If foreign objects are but there is no signal from objects in the battery present, they must be the LED lamp. compartment. Ensure that removed.
Guidelines and manufacturer's declaration – electromagnetic emissions The model ASG 341 is intended for use in an environment as specified below. The customer or the user of the model ASG 341 should ensure that it is used in such an environment. Electromagnetic Compliance Electromagnetic environment –...
Page 16
Guidelines and manufacturer's declaration – electromagnetic immunity The model ASG 341 is intended for operation in an electromagnetic environment as specified below. The customer or the user of the model ASG 341 should ensure that it is used in such an environment.
Page 17
12.0 Information regarding electromagnetic immunity Guidelines and manufacturer's declaration – electromagnetic immunity The model ASG 341 is intended for operation in an electromagnetic environment as specified below. The customer or the user of the model should ensure that it is used in such an environment.
ASG 341 must be observed at every application site. If unusual performance characteristics are observed, additional measures might have to be taken, such as reorientation or relocation of the model ASG 341. b. Not applicable above the frequency range from 150 kHz to 80 Mhz.
Serial number, 00001 (consecutive number) Date of manufacture, 2013-04 (year, month) 2013-04 The snore stopper ASG 341 is certified in accordance with the EU directive 93/42 EEC concerning medical devices. 0123 Manufacturer: Handelshaus Dittmann GmbH, Kissinger Straße 68, D-97727 Fuchsstadt/Germany Protection against electric shock according to type BF (body float).
The statutory warranty period is 24 months as of the purchase date for material and manufacturing defects of the product. Please keep the receipt as proof of the purchase of the snore stopper ASG 341 in order to be able to assert a possible warranty claim. The warranty does not include:...
Need help?
Do you have a question about the ASG 341 and is the answer not in the manual?
Questions and answers