Front Panel Connector Repair
The front panel assembly is an exchange assembly. However, if front panel
sensor cables or power reference cables are faulty they should be replaced
by ordering the appropriate kit and following the assembly/disassembly
procedure. The instrument accuracy test should then be carried out to
verify the functionality of the new part.