STEP1 B
If there is an empty insulin cartridge, remove it
STEP1 C
1
If the dose window does not show the number "0",
push the injection button completely down
2
Hold the plunger reset ring
3
Turn the pen body towards you until it cannot be
turned further and the plunger is fully retracted
IMPORTANT
- Do not touch the disk at the tip of the plunger and do not try to push the
plunger back in
- This could damage the plunger and cause malfunction of the pen
Unscrew
the
pen body
from the
Fully rewind
the starting position
cartridge holder
plunger
the
0
0
1
1
to
11