To Remove The Power Supply; Figure 36 Removing The Power Supply - Agilent Technologies 5000 Series Service Manual

Hide thumbs Also See for 5000 Series:
Table of Contents

Advertisement

To remove the power supply

Power Supply Cable
5000 Series Oscilloscopes Service Guide
1 Perform the previous procedures:
"To remove the cabinet"
"To remove the main shield"
"To remove the AC input board"
2 Remove the 4 T10 screws securing the power supply to the
deck. then remove the power supply.
3 Disconnect the power supply and AC cables
AC Cable
Figure 36
Removing the power supply
4 Reverse this procedure to install the power supply.
Replacing Assemblies
on page 80
on page 89
on page 103
Power Supply
T10 Screws (4)
5
105