To Check The System Board; To Check The Display - Agilent Technologies 5000 Series Service Manual

Hide thumbs Also See for 5000 Series:
Table of Contents

Advertisement

To check the system board

To check the display

5000 Series Oscilloscopes Service Guide
4 If any of the test point voltages are not within the specified
range, replace the AC input board.
1 Disconnect J2730 from the system board and measure pins
19 and 20 on the J2730 connector. This is the voltage to the
display, and it should be +3.3 V. If it is not, replace the system
board. If the voltage is correct, replace the display.
2 Disconnect J2750 from the system board and measure pins 3
and 4 on the J2750 connector. This is the voltage to the
inverter, and it should be +5 V. If it is not, replace the system
board. If the voltage is correct, replace the display.
3 If all cables are properly connected and none of the previous
tests confirm a failure on another assembly, replace the
system board.
1 Disconnect the power cord.
2 Check to verify that the backlight inverter cable is connected.
3 Ensure the display LCD cable is connected.
4 Connect the power cord.
5 Use the DMM to check the Inverter Power voltage (see table
below).
6 If the voltage is incorrect, replace the system board.
7 If the voltage is correct, use an oscilloscope to check the LCD
clock (see table below).
Troubleshooting
4
71