...4.4.1 Setup
English
Language
None
Global Alarm Ack Source
SM1000
Instrument tag

4.4.2 Security

Security
Setup
Logs
Operator Messages
User 1
Operator 1
Operator 2
User 2
User 2
Name
Password
Access
Operator 3, No access
User 3
Operator 4, No access
User 4
Password protected
Configuration security
Setup level security
On
Setup Level Password
****
Operator 2
****
Enabled
OK
Select the language to be used to display standard user
prompts and menu items.
A digital source can be used to acknowledge active alarms in
both Process Groups 1 and 2. Refer to APPENDIX 1 for a full
list of sources available.
Enter the tag to be used to identify the instrument on
configuration and audit log files.
Change the user name and password used to access the
configuration level when 'Password protection' is enabled –
see below.
Note. User 1 has 'administrator' type priviledges
allowing them to change any of the user names and
passwords. Other users can change only their own user
name and password. For users 2, 3 and 4 access can be
enabled or disabled. If it is disabled then the configuration
levels cannot be accessed using that user name.
When a change to the instrument's configuration occurs, the
user's ID is added to the audit log entry.
Set the method of access to the Configuration level. If
'Password protected' is selected, access is by means of the
password set above.
Note. If 'Switch protected' is selected, access to the
Configuration level for all users is prohibited once the
changes have been saved and made active. Access to the
configuration level can then be achieved only by setting the
security switch to the 'Configuration not protected' position
– see Fig. 4.1.
When set to 'On', access to the Setup level is protected by a
single password for all users.
Displayed only when Setup Level Security is set to 'On'
4 CONFIGURATION...