Emc Information - DeVilbiss DV6WM Instruction Manual

Table of Contents

Advertisement

Available languages

Available languages

EMC INFORMATION

(per IEC 60601-1-2:2007 Section 5.2.2.2)
Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
The DeVilbiss DV6WM Wireless Modem is intended for use in the electromagnetic
environment specified below. The customer or the user of the DeVilbiss DV6WM
Wireless Modem should assure that it is used in such an environment.
IEC 60601
Compliance
Immunity Test
test level
level
V1 = 3 Vrms
3 Vrms
Conducted RF
6 Vrms on ISM
150 kHz to
IEC 61000-4-6
& Amateur
80 MHz
Bands
Radiated RF
10 V/m
E1 = 10 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the DeVilbiss DV6WM
Wireless Modem is used exceeds the applicable RF compliance level above, the DeVilbiss
DV6WM Wireless Modem should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or
relocating the DeVilbiss DV6WM Wireless Modem.
b.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
EN - 12
Electromagnetic Environment – Guidance
Portable and mobile RF communications
equipment should be separated from the
device by no less than the recommended
separation distances calculated/listed below:
D=(0.4)√ P
D=(0.4)√ P 80 to 800 MHz
D=(0.7)√ P 800 MHz to 2.5 GHz
Where P is the maximum power rating in
watts and D is the recommended separation
distance in meters.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey
, should be less than the compliance
a
level in each frequency range
.Interference
b
may occur in the vicinity of equipment marked
with the following symbol:
(per IEC 60601-1-2:2007 Section 5.2.2.2)
Recommended Separation Distances Between Portable and
Mobile RF Communications Equipment and this device. This
device and system are NOT Life-Supporting
The DeVilbiss DV6WM Wireless Modem is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user of
the DeVilbiss DV6WM Wireless Modem can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the DeVilbiss DV6WM Wireless Modem as recommended
below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter (m)
Rated Maximum
150 kHz to 80 MHz
Output Power of
outside ISM bands
Transmitter (W)
D=(0.4)√ P
0.01
0.04
0.1
0.11
1
0.35
10
1.1
100
3.5
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz
are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and
40.66 MHz to 40.70 MHz.
NOTE 3: An additional factor of 10/3 has been incorporated into the formulae used in
calculating the recommended separation distance for transmitters in the ISM frequency
bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to
decrease the likelihood that mobile/portable communications equipment could cause
interference if it is inadvertently brought into patient areas.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
80 MHz to 800 MHz
800 MHz to 2.5 GHz
D=(0.4)√ P
D=(0.7)√ P
0.04
0.07
0.11
0.22
0.35
0.70
1.1
2.2
3.5
7.0
SE-DV6WM-1

Advertisement

Table of Contents
loading

Table of Contents