3. Calibration

The OME-A-8111 is calibrated to its best state of operation. For environments with large
vibration, recalibration is recommended. Before calibrating the OME-A-8111, users should
have the following items:
One 6-digit multi-meter.
One stable voltage source (4.9988V)
Diagnostic program: this program is included in the delivered package
and will guide the user to in proceeding with the calibration.

3.1 Calibration VR Description

There are seven VRs on the OME-A-8111. Calibration is needed to adjust all seven VRs.
VR Num. Description
VR1
D/A Gain adjustment
VR2
D/A Offset adjustment
VR3
A/D Offset adjustment
VR4
A/D's Gain adjustment
VR5
A/D's PGA(Programmable Gain Amplifier) Offset Adjustment
VR4
A/D Converter
VR3
D/A Converter
OME-A-8111 Hardware Manual (ver.1.1, Jul/2003)
PGA
VR5
VR1
8 ch.
Mux
VR2
O.P.
CN1
29