Fluke 79 III Calibration Information Manual page 4

Series iii multimeter
Table of Contents

Advertisement

79/26
Calibration Information
Accuracy specifications are given as: ± ([% of reading] + number of least significant digits) ac readings
are ac-coupled, true rms, and are valid from 5 % to 100 % of range. Add ±(2 % of reading + 2 % of
range) for 1.4 < CF ≤ maximum CF. Maximum CF is 3 full scale, 6 half scale.
Function
K
(45 Hz to 1 kHz)
L
L*
m
e
Capacitance
R
40Ω
(Lo-Ohms)
G
In 40 Ω and 40 mV ranges, thermals may introduce additional errors. Maximum accuracy is obtained when
*
both probe tips are maintained at the same temperature.
**
Accuracy applies when measuring film capacitors or better and the open lead reading is subtracted from the
measurement. This meter uses a dc-type measurement technique.
*** Accuracy applies after lead resistance compensation.
4
Table 1. Specifications (cont.)
Range
Resolution
400.0 mV
0.1 mV
4.000 V
0.001 V
40.00 V
0.01 V
400.0 V
0.1 V
1000 V
1 V
4.000 V
0.001 V
40.00 V
0.01 V
400.0 V
0.1 V
1000 V
1 V
40.00 mV
0.01 mV
400.0 mV
0.1 mV
400.0 Ω
0.1 Ω
4.000 kΩ
0.001 kΩ
40.00 kΩ
0.01 kΩ
400.0 kΩ
0.1 kΩ
4.000 MΩ
0.001 MΩ
40.00 MΩ
0.01 MΩ
99.99 nF
0.01 nF
999.9 nF
0.1 nF
9.999 µF
0.001 µF
99.99 µF
0.01 µF
999.9 µF
0.1 µF
9999 µF
1 µF
400 Ω
0.1Ω
40 Ω*
0.01 Ω
400 Ω
0.1 Ω
2.450 V
0.001 V
Accuracy
Burden Voltage (Typical)
±(1.9 %+4)
±(1.9 %+2)
±(1.5 %+2)
±(1.5 %+2)
±(1.5 %+2)
±(0.3 %+1)
±(0.3 %+1)
±(0.3 %+1)
±(0.3 %+1)
±(0.3 %+5)
±(0.3 %+1)
±(0.4 %+2)
±(0.4 %+1)
±(0.4 %+1)
±(0.6 %+1)
±(0.4 %+1)
±(1 %+3)
±(1.9 %+2)**
±(1.9 %+2)**
±(1.9 %+2)**
±(1.9 %+2)**
±(1.9 %+2)**
±10 % Typical
5 % Typical***
5 % Typical***
5 % Typical***
±2 % Typical
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable