Mastech MAS838 Operator's Instruction Manual

Mastech MAS838 Operator's Instruction Manual

Hide thumbs Also See for MAS838:

Advertisement

Mastech MAS838 Operator's Instruction Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MAS838 and is the answer not in the manual?

Questions and answers

Summary of Contents for Mastech MAS838

  • Page 1 Distributed by: www.Jameco.com 1-800-831-4242 The content and copyrights of the attached material are the property of its owner. Jameco Part Number 220741...
  • Page 2: Digital Multimeter

    OPERATOR’S INSTRUCTION MANUAL DIGITAL MULTIMETER...
  • Page 3: Safety Information

    SAFETY INFORMATION This multimeter has been designed according to IEC −1010 concerning electronic measuring instruments with overvoltage category (CATⅡ) and pollution 2. Follow all safety and operating instructions to ensure that the meter is used safely and is kept in good operating condition. Full compliance with safety standards can be guaranteed only with test leads supplied.
  • Page 4: During Use

    DURING USE • Never exceed the protection limit values indicated in specifications for each range of measurement. • When the meter is linked to measurement circuit, do not touch unused terminals. • Never use the meter to measure voltages that might exceed 600V above earth ground in category Ⅱ...
  • Page 5: Front Panel

    FRONT PANEL...
  • Page 6: Front Panel Description

    FRONT PANEL DESCRIPTION Display 3 1/2 digit, 7 segment, 15mm high LCD. Rotary switch This switch is used to select functions and desired ranges as well as to turn on/off the meter. Hold button When this button is pushed, the display will keep the last reading and “...
  • Page 7: Diode & Continuity

    Low battery indication :“ ”appears on the display Size :138mm×69mm×31mm Weight : Approx.170g. DC VOLTAGE Range Resolution Accuracy ±0.5% of rdg ± 2 digits 200mV 100μV ±0.5% of rdg ± 2 digits ±0.5% of rdg ± 2 digits 10mV ±0.5% of rdg ± 2 digits 200V 100mV ±0.8% of rdg ±...
  • Page 8: Transistor Hfe Test

    TRANSISTOR hFE TEST (0-1000) Range Test Range Test Current Test Voltage NPN & PNP 0-1000 Ib=10μA Vce=3V RESISTANCE Range Resolution Accuracy ±0.8% of rdg ± 3 digits 200Ω 0.1Ω ±0.8% of rdg ± 2 digits 2kΩ 1Ω ±0.8% of rdg ± 2 digits 20kΩ...
  • Page 9: Dc Current Measurement

    DC CURRENT MEASUREMENT Connect the red test lead to the “V.Ω.mA” jack and the black test lead to “COM” jack. (For measurements between 200mA and 10A, remove red lead to“10A” jack.) Set the rotary switch at desired DCA position. Open the circuit in which the current is to be measured, and connect test leads in series with the circuit.
  • Page 10: Resistance Measurement

    RESISTANCE MEASUREMENT Connect the red test lead to “V. Ω. mA” jack and black test lead to the “COM” jack. (The polarity of red lead is positive “+”.) Set the rotary switch at desired ”Ω” range position. Connect test leads across the resistor to be measured and read LCD display.
  • Page 11 in the bottom of the case. Simply remove the old, and replace with a new one. Be careful to observe battery polarity. WARNING Before attempting to open the case, always be sure that test leads have been disconnected from measurement circuits. Close case and tighten screws completely before using the meter to avoid electrical shock hazard.