10. Troubleshooting

Problem/Ques-
Possible Cause/Remedy
tion
The atomiser
1. Too much or too little medicine in
produces
the atomiser.
little or no
Minimum: 0.5 ml, Maximum: 6 ml.
aerosol.
2. Check that the nozzles are not
blocked If necessary, clean the
nozzle (e.g. by rinsing). Use the
atomiser again.
CAUTION: Only poke implements
through the fine holes from the under-
side of the nozzle.
3. The atomiser is not held in an
upright position.
4. The medicine solution is unsuitable
for nebulising (e.g. too thick). The
medicine solution should be speci-
fied by the doctor.
Output is too
Hose kinked, filter blocked, too much
low.
inhalation solution.
Which
Consult your doctor.
medicines are
Essentially all medicines that are suit-
suitable for inha-
able and approved for device-based
lation?
inhalation can be inhaled.
Some inhalation
This is normal and occurs for techni-
solution
cal reasons. Stop inhalation if you
remains in the
hear a notable difference in the noise
atomiser.
of the atomiser.
What special
1. In babies, the mask should cover
steps should be
the mouth and nose to guarantee
taken for babies
effective inhalation.
and children?
2. In children, the mask should also
cover the mouth and nose. Nebu-
lisation next to a sleeping person
is not suitable because insufficient
medicine reaches the lungs.
Note: Children should only use the
device with help and under supervi-
sion of an adult. Never leave a child
alone with the nebuliser.
Inhalation with
This is for technical reasons. Less
the mask takes
medicine is inhaled per breath
longer.
through the mask holes than using
the mouthpiece. The aerosol is mixed
with ambient air through the holes in
the mask.
Problem/Ques-
Possible Cause/Remedy
tion
Why do I need
There are two reasons for this:
to replace the
1. To guarantee a therapeutically
atomiser at
regular inter-
vals?
2. You are also recommended to
Does each per-
This is absolutely necessary for
son need their
hygiene reasons.
own atomiser?

11. Technical specifications

Dimensions
295 x 178 x 94 mm
(WxHxD)
1,6 kg
Weight
Operating
approx. 0,64 - 0,96 bar
pressure
Atomiser fill
max. 6 ml
volume
Medicine flow
> 0.2 ml/min.
Sound pressure
< 60 dBA
(acc. to DIN EN 13544-1 section 26)
Mains
connection
230 V ~, 50 Hz; 180 VA
Operating
Temperature: +10 °C – +40 °C
conditions
Relative humidity: 10% – 95%
Storage and
Temperature: -21 °C – +70 °C
transport
Relative humidity: 10% – 95%
conditions
Atmospheric pressure: 500 – 1060 hPa
Subject to technical modifications.
Particle size diagram
17
effective particle spectrum, the
nozzle holes should not exceed
a certain diameter. Due to the
thermal and mechanical stresses,
the plastic is subjected to a certain
amount of wear. The atomiser [9] is
particularly sensitive. This can also
cause changes to the droplet com-
position of the aerosol, which has a
direct effect on the effectiveness of
the treatment.
replace the atomiser on a regular
basis for hygiene reasons.
Particle size [μm]