Important Information
5. Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The ElectroVita is intended for use in the electromagnetic environment specified below. The
customer or the user of the ElectroVita should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
NOTE: UT is the AC mains voltage prior to application of the test level
6. Guidance and manufacturer's declaration – electromagnetic immunity
The ElectroVita is intended for use in the electromagnetic environment specified below. The
customer or the user of the ElectroVita should assure that it is used in such an environment.
19
IEC 60601
Compliance
Test Level
Level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1kV for
±1kV for
input/output
input/output
lines
lines
±1 kV line(s)
±1 kV line(s)
and neutral
and neutral
<5 % UT
<5 % UT
(>95 % dip in UT)
(>95 % dip in UT) commercial or hospital environment. If a dips or an
for 0,5 cycle
for 0,5 cycle
40 % UT
40 % UT
(60 % dip in UT)
(60 % dip in UT)
for 5 cycles
for 5 cycles
70 % UT
70 % UT
(30 % dip in UT)
(30 % dip in UT)
for 25 cycles
for 25 cycles
<5 % UT
<5 % UT
(>95 % dip in UT)
(>95 % dip in UT)
for 5s
for 5s
3 A/m
Not applicable
Electromagnetic Environment Guidance
Floors should be wood, concrete or ceramic tile.
If oors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
interruption of mains power occurs, the current of the
ElectroFlex may be dropped o from normal level, it
may be necessary to use uninterruptable power
supply or a battery
Not applicable
Need help?
Do you have a question about the circulation enhancer and is the answer not in the manual?
Questions and answers
Does this need rechargeable battery's to function
The JML Circulation Enhancer can operate using a DC adapter connected to a mains outlet or with three 1.5V alkaline type C batteries. It does not specifically require rechargeable batteries.
This answer is automatically generated
My machine is not displaying numbers