Troubleshooting and Maintenance
SYMPTOM
CAUSE
Device will
Batteries inserted in wrong direction
not turn on
The adaptor is not connected to the
device correctly
Power turns o
Gel pads not attached
too soon
correctly to the skin
Power turns o
If you are using the batteries operation, Fit three new 1.5V alkaline batteries type C
while using
then the batteries are weak/exhausted
massager
Treatment period of 25 minutes is over Restart treatment or turn o the massager
and power turns o automatically
If you are using the body massager,
the cable may be damaged
It is di cult to
Transparent lm not peeled o
attach gel pad
Gel pad applied immediately
to the skin
after washing
Adhesive surface of gel pad damaged
The gel pads get dirty and lose their
adhesive/stickiness
Adhesive surface
Use of gel pad during perspiring
of gel pad is sticky Gel pad washed too long and/or
too frequently
Gel pads stored under high temperature,
high humidity or direct sunshine
It is di cult to
The soles of your feet are too dry
feel stimulation
Your soles are not placed on the foot
plates correctly
Gel pads not attached correctly to
the skin
Gel pads overlap each other
Electrode cord not connected correctly Connect electrode cord correctly
Applied intensity too weak
The skin turns
Adhesive surface of Gel pads is
red or the skin
dirty or dry
feels irritated
Adhesive surface of Gel pads damaged Replace gel pads
Hygiene
After using the product
Storage
Keep the product clean and dust free store within the following condition.
Storage temperature and humidity -10˚C to 60˚C, 10% to 95% RH
Operating temperature and humidity 10˚C to 40˚C, 30% to 90% RH
ACTION
Insert batteries in correct direction or check the
battery is not discharged
Check the connection of the adaptor jack is properly
connected to the device and in the wall socket
Attach Gel pads correctly to the skin
Replace electrode cord
Peel o
lm on the adhesive surface of gel pad
Allow gel pad to dry
Replace gel pad
Replace gel pad or clean the gel pad with a small drop
of water rubbed onto the sticky side of the electrode pad
Chill the gel pad in a domestic fridge for 3-4 hours
Dampen the sole of your feet this will
increase conductivity
Ensure both of your soles are aligned on each foot
plate correctly
Attach gel pad rmly to the skin
Reattach Long Life pads with no overlap
Increase the intensity by pressing the intensity button
Wash adhesive surface of gel pads carefully whilst
unplugged from machine and leave to dry thoroughly
16
Need help?
Do you have a question about the circulation enhancer and is the answer not in the manual?
Questions and answers
Does this need rechargeable battery's to function
The JML Circulation Enhancer can operate using a DC adapter connected to a mains outlet or with three 1.5V alkaline type C batteries. It does not specifically require rechargeable batteries.
This answer is automatically generated
My machine is not displaying numbers