Range
Accuracy
Frequency Resolution
Distortion
Amplitude
Amplitude Accuracy
Amplitude Stability
Outputs
Channel 1 Output
Offset
Expand
Reference
Channel 1 Output
X, Y Outputs
(rear panel)
Aux. A/D Inputs
Aux. D/A Outputs
Sine Out
TTL Out
Data Buffer
Trigger In (TTL)
Remote Pre-amp
Computer Interface
Power
Dimensions
Weight
Family of Brands – ILX Lightwave
1 MHz to 102 kHz
25 ppm + 30 µHz
4-1/2 digits or 0.1 MHz, whichever is greater
-80 dBc (f<10 kHz), -70 dBc (f> 10 kHz) @ 1 Vrms
amplitude
0.004 to 5 Vrms into 10 kΩ (2 mV resolution), 50 Ω
output impedance, 50 mA maximum current into 50
50 ppm/°C
Sine, TTL
When using an external reference, both outputs are
phase locked to the external reference.
DISPLAYS
±10 V output of X, R, X-noise, Aux 1 or Aux 2.
Updated at 512 Hz.
X, Y, R can be offset up to ±105% of full scale
X, Y, R can be expanded by 10x or 100x
4-1/2 digit LED display
INPUTS AND OUTPUTS
±10 V output of X, R, X-noise, Aux 1 or Aux 2.
Updated at 512 Hz.
In-phase and quadrature components
(±10 V), updated at 256 kHz
4 BNC inputs, ±10 V, 1 mV resolution, sampled at
4 BNC outputs, ±10 V, 1 mV resolution
Internal oscillator analog output
Internal oscillator TTL output
8k point buffer with data recorded at rates to 512 Hz
and read through the computer interfaces.
Trigger synchronizes data recording
Provides power to the preamplifier integrated inside
Oriel's LIDA detectors, using cable Model CBL-
70054-LIDA
IEEE-488.2 and RS-232 interfaces standard. All
instrument functions can be controlled and read
through IEEE-488.2 or RS-232 interfaces. TracQ Basic
software utilizes GPIB communication.
GENERAL
100 to 120 VAC, 220 to 240 VAC, 50 to 60 Hz, 40 W
17" W x 5.25" H x 19.5" L [431 x 133 x 495 mm]
30 lbs [13.6 kg]
• New Focus™ • Ophir
• Corion • Richardson Gratings™ • Spectra-Physics
®
®
MLIDA, Rev A
Ω
1%
512 Hz
®