4.5 A OFF Menu
AUTO SHUT OFF
Set the automatic shut off to 5, 10, 15 minutes (shut off
after 5 minutes, etc.) or OFF (scale stays on).
4.6 FILTR Menu
FILTERING LEVEL
Set the filter level to LO (low), MEd (medium), or HI
(high).
4.7 AZT Menu
AUTO ZERO TRACKING RANGE
Set the automatic zero tracking range to OFF, 0.5d, 1d,
3d (divisions).
4.8 LFT Menu
LEGAL FOR TRADE (LFT) MODE
Set the legal for trade status to ON or OFF.
4.9 END
End menu item.

5. LEGAL FOR TRADE

5.1 Settings
Enter the menu to verify the settings and perform a calibration as explained in
Section 4.2. Set the LFT menu to ON.
NOTE: When LFT is set to ON, the following menu settings cannot be
accessed/changed: Calibration, Resolution, Unit, AZT, LFT.
To regain access to the locked menu settings, remove the small cover on
the bottom of the scale. Then during power up press and hold the
recessed lock switch momentarily until zero is displayed.
EN-9
EN-10
Figure 5-1. Lock Switch.
When accessing the bottom of the scale, avoid placing the scale upside
down on the pan or sub-platform. Place the scale on its side.
5.2 Verification
Before this product can be used in a trade approved application, it must be inspected
in accordance with local weights and measures regulations. It is the responsibility
of the purchaser to ensure that all pertinent legal requirements are met. Please
contact your local weights and measures office for further details.
5.3 Sealing
The weights and measures official can apply a wire or paper security seal as shown
below.
Figure 5-2. Sealing.