Calibration - Health O Meter 500KL User Instructions

Eye level digital beam scale
Hide thumbs Also See for 500KL:
Table of Contents

Advertisement

CALIBRATION

This scale has been factory calibrated, and does not require calibration prior to use. If
required, the scale can be calibrated. Calibration of your 500KL scale is performed using
kilograms (KG) or pounds (LB), according to the units of measure used upon entering into the
calibration path. The following calibration assumes the scale is set to pounds, and 100 lbs (40
kg) / 200 lbs (80 kg) / 300 lbs (120 kg) is employed as the calibration standards.
Press and hold the LB/KG and ZERO buttons at the same time, and press
the ON/OFF button to turn on the scale. "CAL" will briefly appear on the
display, followed by four numbers indicating that you can calibrate. Release
all three buttons.
Press the ZERO button. "C-100" (for LB mode) or "C-40" (for KG mode) will
flash on the display, followed by "0"or "1".
Place 100 lbs or 40 kg on the scale. Four numbers will appear. When
calibration numbers stabilize, press the ZERO button. "C-200" (for LB mode)
or "C-80" (for KG mode) will appear on the display, followed by "0" or "1".
Place 200 lbs or 80 kg on the scale. Four numbers will appear. When
calibration numbers stabilize, press the ZERO button. "C-300" (for LB mode)
or "C-120" (for KG mode) will appear on the display, followed by "0" or "1".
Place 300 lbs or 120 kg on the scale. Four numbers will appear. When
calibration numbers stabilize, press the ZERO button. "END" will appear on
the display, followed by "OFF". The scale will automatically shut off.
 
ACTION
22 
DISPLAY SHOWS
"CAL" followed by
"XXXX"
"XXXX" followed by
"C-100" (LB mode)
or
"C-40" (KG mode)
then "0" or "1".
"XXXX" followed by
"C-200" (LB mode)
or
"C-80" (KG mode)
then "0" or "1".
"XXXX" followed by
"C-300" (LB mode)
or
"C-120" (KG mode)
then "0" or "1".
"XXXX" followed
by "END" then
"OFF"

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents