5-6. CHECKS PRIOR TO REPAIRS
These checks are performed before replacing parts according to
"approximate specifications" to determine the faulty locations. For
details, refer to "Checks Prior to Parts Replacement and
Adjustments" (See page 9).
5-6-1. Temperature Compensation Offset Check
When performing adjustments, set the internal temperature and room
temperature to 22 to 28ºC.
Checking Procedure:
1. Rotate the l L knob to display "MD TEMP CHECK".
2. Press the ENTER/YES button.
3. "MD T=@@(##) [OK]" should be displayed. If "MD T=@@
(##) [NG]" is displayed, it means that the results are bad.
(@@ indicates the current value set, and ## indicates the value
written in the non-volatile memory.)
5-6-2. Laser Power Check
Before checking, check the IOP value of the optical pick-up.
(Refer to 5-8. Recording and Displaying the IOP Information.)
Connection :
Optical pick-up
objective lens
BD board
CN110 pin 5 (I+3V)
CN110 pin 4 (IOP)
Checking Procedure:
1. Set the laser power meter on the objective lens of the optical
pick-up. (When it cannot be set properly, press the m button
or M button to move the optical pick-up.)
Connect the digital volt meter to CN110 pin 5 (I+3V) and
CN110 pin 4 (IOP).
2. Then, rotate the l L knob and display "MD LDPWR
CHECK".
3. Press the ENTER/YES button once and display "MD 0.9 mW
*** $
". Check that the reading of the laser power meter
become 0.84 to 0.92 mW. (*** means IOP value)
4. Press the ENTER/YES button once more and display "MD 7.0
mW *** $
". Check that the reading the laser power meter
and digital volt meter satisfy the specified value.
Specified Value :
Laser power meter reading : 7.0 ± 0.2 mW
Digital voltmeter reading : Optical pick-up displayed value ± 10%
(Optical pick-up label)
(For details of the method for checking
KMS
260A
this value, refer to "5-8. Recording and
27X40
Displaying the IOP Information".)
B0825
Iop = 82.5 mA in this case
Iop (mA) = Digital voltmeter reading (mV)/1 ( Ω )
5. Press the MENU/NO button and display "MD LDPWR
CHECK" and stop the laser emission.
(The MENU/NO button is effective at all times to stop the laser
emission.)
24
Note 1: After step 4, each time the ENTER/YES button is pressed,

5-6-3. Traverse Check

Connection :
Checking Procedure:
1. Connect an oscilloscope to CN110 pin 3 (TE) and CN110 pin
2. Load a disc (any available on the market). (Refer to Note 1.)
3. Press the m button and move the optical pick-up outside the
Laser power
meter
4. Rotate the l L knob and display "MD EF MO CHECK".
5. Press the ENTER/YES button and display "MD EFB =
6. Observe the waveform of the oscilloscope, and check that the
Digital volt meter
7. Press the ENTER/YES button and display "MD EFB =
8. Observe the waveform of the oscilloscope, and check that the
9. Press the ENTER/YES button display "MD EFB =
the display will be switched between "MD LD 0.7 mW $
", "MD LD 0.7 mW $
"MD LD WP
$
Nothing needs to be performed here.
BD board
CN110 pin 3 (TE)
CN110 pin 1 (VC)
1 (VC) of the BD board.
pit.
MO-R".
(Laser power READ power/Focus servo ON/tracking servo OFF/
spindle (S) servo ON)
specified value is satisfied. Do not rotate the l L knob.
(Read power traverse checking)
(Traverse Waveform)
VC
Specified value : Below 10% offset value
IA – BI
Offset value (%) =
X 100
2 (A + B)
MO-W".
specified value is satisfied. Do not rotate the l L knob.
(Write power traverse checking)
(Traverse Waveform)
VC
Specified value : Below 10% offset value
IA – BI
Offset value (%) =
X 100
2 (A + B)
Then, the optical pick-up moves to the pit area automatically
and servo is imposed.
", "MD LD 6.2 mW $
",
", "MD LD 0.9 mW $
".
Oscilloscope
V : 0.5 V/div
H : 10 ms/div
Input : DC mode
A
B
A
B
MO-P".