Fluke 1587 FC User Manual page 16

Hide thumbs Also See for 1587 FC:
Table of Contents

Advertisement

1587 FC/1587/1577
Users Manual
Switch
Position
Temperature from - 40 °C to + 537 °C (- 40 °F to + 998 °F).
k
Celsius is the default temperature measurement unit. The temperature measurement you select is
retained in memory when the Meter is turned off.
o
Ohms from 0.1 Ω to 50 MΩ.
Capacitance from 1 nF to 9999 μF.
Continuity test. Beeper turns on at <25 Ω and turns off at >100 Ω.
X
O
Diode test. There is no ranging in this function. Displays  above 6.600 V.
AC mA from 3.00 mA to 400 mA (600 mA overload for 2 minutes maximum).
c
DC mA from 0.01 mA to 400 mA (600 mA overload for 2 minutes maximum).
Ohms from 0.01 MΩ to 2 GΩ. The last selected output voltage setting is retained in memory when the
Meter is turned off.
Ohms from 0.01 MΩ to 600 MΩ. The last selected output voltage setting is retained in memory when the
Meter is turned off.
Performs insulation test with:
a
50 V, 100 V, 250 V, 500 V (default), and 1000 V source
INSULATION
500 V (default) and 1000 V source
50 V (default) and 100 V source
Press the blue button to activate smoothing during insulation testing.
8
Table 4. Rotary Switch Selections (cont.)
Measurement Function
• • •
• • • •
• • •
• • • •
• • •
• • • •
• • •
• •
• • •