Page 1
1030 WARNING Read and understand these Instructions for Use and all safety precautions it contains. Improper use can cause serious or fatal injury and/or illness. CAUTION––US Federal law restricts this device to sale by or on order of a physician.
Page 2
REGISTER FIRST WARRANTY ACTIVATION REQUIRED Three di erent options to choose from. www.pari.com/register 1.866.335.6943 Registration card enclosed ETL CLASSIFIED Intertek Conforms to UL 60601-1 3045080 Certi ed to CAN/CSA C22.2 No. 601.1 QUESTIONS ABOUT ERAPID? CALL 1-866-335-6943...
TABLE OF CONTENTS INTRODUCTION These Instructions For Use contain information and safety Introduction ................3 precautions for the eRapid ™ Nebulizer System. Save these Safety Precautions And Important Safeguards ..... 4-5 Instructions For Use for future reference. Before using your eRapid Nebulizer System, you should read and Getting Started .................
SAFETY PRECAUTIONS AND IMPORTANT SAFEGUARDS READ ALL DANGERS AND WARNINGS BEFORE USING 4. Always unplug this product immediately after using and before cleaning. DANGER 5. Before use, check your eRapid Nebulizer System or eRapid To reduce the risk of serious or fatal injury from electrocution: Nebulizer Handset for proper assembly.
SAFETY PRECAUTIONS AND IMPORTANT SAFEGUARDS 14. This product contains small parts that may present a WARNING choking hazard to small children. The eRapid Nebulizer 11. To reduce the risk of infection, illness, and/or injury from Connection Cord also presents a strangulation hazard. contamination, clean and dry all parts of the eRapid 15.
GETTING STARTED Check your eRapid Nebulizer System package to ensure you have the following items. Note that two complete Nebulizer Handsets are included. EBASE CONTROLLER NEBULIZER CONNECTION CORD CARRYING CASE AC POWER SUPPLY MEDICATION CAP (NOTE: THE CAP SEAL AND CAP WILL COME AS ONE ASSEMBLED PART) AEROSOL HEAD...
USING “AA” BATTERIES USING AC POWER SUPPLY 1. The eRapid Nebulizer System is designed to be used with 1. The Controller can also be used with the AC Power Supply. The “AA” batteries or with the optional AC Power Supply. Four (4) AC Power Supply is useful as a back-up power supply when the new “AA”...
ERAPID NEBULIZER HANDSET ASSEMBLY Figure WARNING The eRapid Nebulizer Handset is not sterile. Clean and Disinfect the eRapid Nebulizer Handset before the first use. (see Section H for Cleaning and Section J for Disinfecting). Inspect all parts to ensure they are clean and are not visibly damaged.
TAKING A TREATMENT 1. Make sure the eRapid Nebulizer Handset is on a flat, NOTE: eBase Controllers supplied with other eFlow Technology stable surface. devices may not have this software feature. 2. Pour only the medication prescribed by your physician specifically for the eRapid Nebulizer System into the eRapid 11.
Page 11
MESSAGE SCREENS & WARNINGS Figure NOTE: Your Controller has an LCD screen that provides information to you. Think of these screens as “electronic coaching” to ensure proper operation of the eRapid Nebulizer STATUS LIGHT System. Symbols on the screen will “alternate”as shown below. In addition, the Controller status light NEAR THE ON/OFF BUTTON will flash red-green and repetitive alarm beep tones will sound.
ERAPID NEBULIZER HANDSET CLEANING Never use a brush or any other object to clean the Aerosol WARNING Head (Figure 28). THOROUGH RINSING OF BOTH SIDES OF THE AEROSOL HEAD HELPS TO PREVENT CLOGGING AND ENSURE To reduce the risk of infection, illness, and/or injury PROPER OPERATION.
USING THE OPTIONAL EASYCARE EFLOW NEBULIZER HANDSET DISINFECTION EASYCARE (Fig. 30), available separately, is a cleaning aid WARNING designed to maintain the standard performance of the eRapid Aerosol Head. It is important not to use EASYCARE more than 2 To reduce the risk of serious or fatal illness caused by times per week to maintain maximum eRapid Aerosol Head life.
FUNCTIONALITY TEST Answer: If you are using the AC Power Supply, check the A Functionality Test takes only a few minutes and is helpful if you adapter connection to be sure it is firmly connected to the suspect the eRapid Nebulizer Handset is not working properly. adapter port located underneath the battery compartment.