Omron BP785N User Manual page 33

Hide thumbs Also See for BP785N:
Table of Contents

Advertisement

HEM-7321-Z_C_M.book Page 30 Monday, June 16, 2014 12:09 PM
9. Guidance and Manufacturer's Declaration
OMRON Automatic Blood Pressure Monitor
Information for accompanying documents in the scope of IEC60601-1-2:2007
Model: BP785N including AC-adapter
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC's and mobile (cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and
create a potentially unsafe situation. Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the IEC60601-
1-2 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels
of electromagnetic emissions for medical devices.
Medical devices manufactured by OMRON Healthcare conform to this IEC60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• The use of accessories and cables other than those specified by OMRON, with the exception of cables sold by OMRON as replacement
parts for internal components, may result in increased emission or decreased immunity of the device.
• The medical devices should not be used adjacent to or stacked with other equipment.
In case adjacent or stacked use is necessary, the medical device should be observed to verify normal operation in the configuration in which
it will be used.
• Refer to further guidance below regarding the EMC environment in which the device should be used.
• The MEDICAL ELECTRICAL EQUIPMENT BP785N including AC-adapter needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in this documentations.
• The Essential Performance of the BP785N including AC-adapter is to measure a blood pressure and a pulse rate and using the memory
function.
The BP785N including AC-adapter may be interfered with by other equipment, even if that other equipment complies with CISPR EMISSION
requirements.