high results of uric acid test.
● In situations of decreased peripheral blood flow, examples would include but
are not limited to severe dehydration, in shock, or in a hyperosmolar state (with
or without ketosis), hypertension, the test results may be falsely low.
● BeneCheck PLUS Uric Acid Test Strips are plasma calibrated, test with serum or
plasma sample will give falsely high results.
● Lipemic samples : Cholesterol levels up to 400 mg/dL (10.35 mmol/L) do not
affect the results. Triglycerides and grossly lipemic patient samples have not
been tested and are not recommended for testing with the BeneCheck PLUS
Uric Acid Test Strips.