Specifications 
Display   
Ranges and Resolution 
Unit of measure 
 
Accuracy   
Sampling time
Tare Control
Transducer type 
 
User Calibration 
 
Data output
Over range indication 
Under range indication 
Low battery indication 
Power supply 
Power consumption 
Operating Temperature 
Operating Humidity 
Dimensions
Weight   
actoolsupply.com
5‐digit multi‐function LCD 
Digit height: 20.9mm (0.8") 
0.10 to 600g (resolution: 0.01g) 
0.005 to 9.9995 oz. (resolution: 0.0005 oz.) 
10.000 to 21.164 oz. (resolution: 0.001 oz.) 
Selectable via rear switch (grams/ounces) 
± (0.05% + 0.04g); within 2 hours of calibration 
Accuracy  applies  at  23
following  RF  environmental  conditions:  field  strength  <  3V/M  and 
frequency < 30Hz only) 
Approx. 1 measurement per second 
600g maximum (approx.) 
Load cell 
Using 200g or 400g standard weights (not supplied) 
RS232 PC interface jack on rear of instrument 
Optional PC interface cable and software PN 407001 
Optional USB adaptor PN 407001‐USB 
Upper dashes ‐ ‐ ‐ ‐ ‐ ‐ appear on the LCD 
Lower dashes _ _ _ _ appear on the LCD 
'LO' Battery icon appears on the LCD 
6  x  1.5V  AA  batteries  or  power  adaptor  9V/300mA  (supplied  with 
multi‐plugs)  
17mA DC approximately 
o
o
0 to 50
C (32 to 122
F) 
< 80% RH 
Base: 250 x 190 x 70mm (9.8 x 7.5 x 2.8") 
Platform diameter:  120 mm (4.7") 
1 kg (2.2 lbs.) 
actoolsupply.com
o
o
C  ±5
C  (specifications  tested  under  the 
9      
  SC600‐EU‐EN v1.1   1/14 