Counter Operation 
For a collection of objects or pieces (each piece having the same weight as the next), the SC600 
can  calculate  the  number  of  these  objects  in  a  collection.  A  group  of  pieces  can  be  used  as  a 
sample example in groups of 10, 20, 50, or 100 pieces. 
1.
Prepare a group of pieces (in a group size of 10, 20, 50, or 100 pieces). This will be the 
'standard'  that  the  SC600  uses  to  count  subsequent  collections  of  pieces.  Do  not  place 
any pieces on the scale yet. 
2.
With the SC600 powered OFF, press and hold the CAL button while switching the SC600 
ON 
3.
Hold the CAL button until one beep is heard 
4.
The display will count down to zero and the bottom of the display will show the COUNT 
and PCS (pieces) icons
5.
The  display  will  scroll  through  the  group  sizes  (10,  20,  50,  and  100)  one  group  size  per 
second 
6.
Place the group of pieces (from step 1) onto the weight platform
7.
Press the CAL button when the appropriate group size (10, 20, 50, and 100) is displaying.
The display will flash twice and freeze the group size number.
8.
Remove the group of pieces from the platform. The SC600 is now ready for counting.
9.
Place an unknown number of pieces (of the same weight as the ones from step 1) and
the SC60 will display the number of pieces presently on the platform.
Change the Count Sample Number 
1. Remove all items from the SC600 platform
2. Press  the  CAL  button  once;  the  display  will  scroll  the  piece  group  sizes  (10,  20,  50,  and
100) once per second
3. Place  a  group  of  pieces  (10,  20,  50,  or  100)  onto  the  weight  platform  and  when  the 
number on the SC600 display matches the number of pieces on the platform press the CAL
button
4. The display will flash twice and freeze the count number 
5. Remove the group of pieces from the platform. The SC600 is now ready for counting.
6. Place an unknown number of pieces (of the same weight as the ones from step 3) and the
SC60 will display the number of pieces presently on the platform.
Counter Operation Considerations 
The one piece weight should be larger than 0.1g otherwise errors may occur in counting. The 
accuracy of the counter depends on the number of units counted. It is strongly recommended 
that more samples, for example 50 units or more, be counted for the best accuracy. 