Data Format - A&D EJ-123 Instruction Manual

Compact precision balance
Hide thumbs Also See for EJ-123:
Table of Contents

Advertisement

Pin Connections
Inside of the balance
The interface is designated as DCE (Data Communication Equipment).
9.3.

Data Format

S
Header
There are four types of headers:
ST :
Stable weighing data (including % data)
QT :
Stable counting data
US :
Unstable weighing data (including counting and % data)
OL :
Out of weighing range (Over)
The data is normally 9 digits including a decimal point and a sign.
There are 10 types of units:
g
: Weighing data "gram"
P C
: Counting data "pcs"
%
: Percentage data "%"
o z
: Weighing data "decimal ounce"
o z t
: Weighing data "troy ounce"
c t
: Weighing data "carat"
d w t
: Weighing data "penny weight"
GN
: Weighing data "grain"
N
: Force data "Newton"
D S
: Calculated density (specific gravity) value
The terminator is always C
Example of output data:
Weighing data "gram"
Counting data
Percentage data
Out of range "gram" (+)
Out of range "pcs" (-)
,
T
+
0
0
Separator
L
.
R
F
S T , + 0 0 1 2 . 3 4 5
Q T , + 0 0 0 1 2 3 4 5
S T , + 0 0 0 1 2 3 . 4
O L , + 9 9 9 9 . 9 9 9
O L ,
D-Sub 9 pin male connector
1
N.C.
2
Transmit data
3
Receive data
4
N.C.
5
Signal ground
6
Data set ready
7
Request to send
8
Clear to send
9
N.C.
.
0
0
0
0
Data
- 9 9 9 9 9 9 9 9
25
0
g C
R
Unit
Terminator
(" " shows a space.)
P C C
P C C
L
F
g C
L
R
F
L
R
F
% C
L
R
F
g C
L
R
F
L
R
F

Hide quick links: