Description Of The Item "Data Format - A&D HR-300i Instruction Manual

Hr-i series analytical balances
Table of Contents

Advertisement

Description of the Item "Data Format"

10.5.
A&D standard format
This format is used when the peripheral equipment can receive the A&D format.
If an AD-8121B is used, set the printer to MODE 1 or 2.
This format consists of fifteen characters excluding the terminator.
A header of two characters indicates the balance condition.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is applied.
The unit, consisting of three characters, follows the data.
Header
Stable header
Unstable header
Overload header
DP (Dump print) format
This format is used when the peripheral equipment can not receive the A&D format.
If an AD-8121B is used, set the printer to MODE 3.
This format consists of sixteen characters excluding the terminator.
A header of two characters indicates the balance condition. No overload header is used.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data
is not zero or overloaded.
The unit, consisting of three characters, follows the data.
Header
Stable header
Unstable header
KF format
This is the Karl-Fischer moisture meter format and is used when the peripheral equipment
can only communicate using this format.
This format consists of fourteen characters excluding the terminator.
This format has no header characters.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data
is not zero or overloaded.
This format outputs the unit only for a stable value.
Data
Data
Data
Unit
Unit
Terminator
Stable header of counting mode
Unit
Terminator
Stable header of counting mode
Terminator
Stable value
Unstable value
32
5if type 0
5if type 1
5if type 2