Data Format - A&D EJ-123 Instruction Manual

Compact precision balance
Hide thumbs Also See for EJ-123:
Table of Contents

Advertisement

Pin Connections
Inside of the balance
The interface is designated as DCE (Data Communication Equipment).
9.3.

Data Format

S
Header
There are four types of headers:
ST :
Stable weighing data (including % data)
QT :
Stable counting data
US :
Unstable weighing data (including counting and % data)
OL :
Out of weighing range (Over)
The data is normally 9 digits including a decimal point and a sign.
There are 14 types of units:
g
: Weighing data "g"
P C : Counting data "pcs"
%
: Percentage data "%"
o z
: Weighing data "decimal ounce"
l b
: Weighing data "decimal pound"
o z t
: Weighing data "troy ounce"
c t
: Weighing data "carat"
m o m
: Weighing data "momme"
d w t
: Weighing data "penny weight"
GN
: Weighing data "grain"
N
: Force data "Newton"
t l
: Weighing data "tael"
t
: Weighing data "tola"
D S
: Calculated density (specific gravity) value
The terminator is always C
,
T
+
0
0
Separator
L
.
R
F
D-Sub 9-pin male connector
1
N.C.
2
Transmit data
3
Receive data
4
N.C.
5
Signal ground
6
Data set ready
7
Request to send
8
Clear to send
9
N.C.
.
0
0
0
0
Data
31
0
g C
R
Unit
Terminator
(" " : Space)
L
F