Cleaning The Outside Of The Unit; Cleaning The Test Strip Holder And Lens - Lifescan SureStep Pro Operator's Manual

Bedside unit professional blood glucose management system
Table of Contents

Advertisement

Care and Maintenance

Cleaning the Outside of the Unit

Clean the outside of the unit with a cloth
moistened with a 10% bleach solution. Follow
with a cloth moistened with water to remove
residual bleach. Dry the unit thoroughly.
CAUTION
Do not get water inside the unit. Never im-
merse the unit or hold it under running water
because it will damage the unit.

Cleaning the Test Strip Holder and Lens

Clean the test strip holder (cover and base),
lens area, and contact points. Use a 10% bleach
solution followed by water.
CAUTION
Do not use alcohol, glass cleaners, or any
cleansers containing abrasives, phenol, or
ammonia to clean the test strip holder or lens
area because it will damage the parts.
Press down on the left side of the test strip
1
holder (the end closest to the power button).
This releases the holder, allowing you to
slide it from the unit.
54