ALARM TEMPERATURE SETTING

This unit is provided with the high and low temperature alarm and the temperature at which the alarm is
activated is changeable.
The setting range of alarm temperature is between the chamber temperature ±5
The following procedure shows the setting of alarm temperature according to the condition below:
High temperature alarm: activates at the temperature 5
Low temperature alarm: activates at the temperature 5
Note: The alarm temperature is set 10
Table 3. Procedure for setting high temperature alarm
Description of operation
1
Press numerical value shift key for
2
about 5 seconds.
Press numerical value shift key
3
and scroll the figure to 1.
4
Press set key.
Scroll the figure to 005 by using
5
digit shift key and numerical value
shift key
6
Press set key.
Note: The alarm temperature is set 10
Table 4. Procedure for setting low temperature alarm
Description of operation
1
Press numerical value shift key for
2
about 5 seconds.
Press numerical value shift key
3
and scroll the figure to 2.
4
Press set key.
Scroll the figure to -05 by using
5
digit shift key and numerical value
shift key
6
Press set key.
• The alarm temperature set mode returns to the temperature display mode automatically when 90
seconds has passed without any key operation.
o
C higher than the set temperature
o
C lower than the set temperature
o
C higher than the set temperature at the factory.
Key operated
----
The current chamber temperature is
displayed.
The first digit is flashed.
When pressed, the figure of settable
digit changes.
The current setting is displayed and
SET
the first digit is flashed.
When pressed, the figure of settable
digit changes.
When pressed, the changeable digit
is shifted.
Alarm temperature is memorized and
SET
the current chamber temperature is
displayed.
o
C lower than the set temperature at the factory.
Key operated
----
The current chamber temperature is
displayed.
The first digit is flashed.
When pressed, the figure of settable
digit changes.
The current setting is displayed and
SET
the first digit is flashed.
When pressed, the figure of settable
digit changes.
When pressed, the changeable digit
is shifted.
Alarm temperature is memorized and
SET
the current chamber temperature is
displayed.
16
o
o
C and ±40
C.
Indication after operation