AirSep FreeStyle Patient Manual page 42

Portable oxygen concentrator
Hide thumbs Also See for FreeStyle:
Table of Contents

Advertisement

FreeStyle™
Problem
FreeStyle does not
operate when a pulse
flow selection button is
pressed.
A continuous alarm
sounds and the (red)
alarm light illuminates
continuously.
Intermittent alarm
condition, and the
(yellow) BATT light
illuminates intermittently.
Rapid alarm condition,
and the (yellow) BATT
light illuminates
intermittently and
FreeStyle shuts down.
®
AirSep
Corporation
MN131-1 rev. B - 02/13
Probable Cause
Battery is discharged.
Malfunction.
FreeStyle has not
detected a breath for
15 minutes.
Battery requires
charging.
Battery voltage is too
low to operate the
FreeStyle unit.
Solution
Power the unit through the
DC outlet, or an AC outlet.
Contact your Equipment
Provider, and change to
another source of oxygen
as necessary.
Check the cannula
connection.
Ensure that cannula
tubing is not kinked.
Make sure that the
cannula is positioned
properly and that you are
breathing through your
nose. For mouth-breathing
your clinician may
recommend a chin strap.
Connect to a DC or an AC
outlet within 2 minutes, or
connect to the optional
AirBelt.
Connect to DC or an AC
outlet immediately.
EN
EN-39

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for AirSep FreeStyle

Table of Contents