10.2.5.3 Removing the batteries

CAUTION: Risk of fire and burns if battery poles are bridged!
• When replacing the batteries the battery poles MUST NOT come into contact with metal parts
of the wheelchair causing bridging.
• Be sure to replace the battery pole caps after the batteries have been replaced.
WARNING: Risk of fire and burns due to damage to the battery cables!
• The battery cables and other cables are positioned in a cable duct above the batteries. The
cable duct protects the cables against crushing and other damage. It may not be removed.
Requirements:
• 6mm Allen key
• 5 mm Allen key
• Needle-nosed pliers
PLEASE NOTE:
If your wheelchair is fitted with a lifter the seat unit must be pulled upward in order to access the
batteries. This work should be carried out by at least two people.