If the meter does not display a message after inserting a
test strip:
POSSIBLE CAUSE
Need battery
No battery or battery
incorrectly installed
Test strip inserted incor-
rectly or incompletely
Defective meter or test strip
If the test does not start after applying the sample:
POSSIBLE CAUSE
Insufficient blood sample
Defective test strip
Sample applied after auto-
matic shutoff (two minutes
after last user action)
Defective meter
WHAT TO DO
Replace the battery.
Check that the battery is
correctly installed with the positive
side up.
"+"
Insert the test strip correctly
with the contact bars end first and
facing up. Push it in until it will go no
further.
Call 1 800 227-8862.
WHAT TO DO
Repeat the test with a new
test strip and a larger sample.
Repeat the test with a new
test strip.
Repeat the test with a new test
strip; apply sample only when
R
the
appears on the display.
Call 1 800 227-8862.