Invacare Comet Service Manual page 45

Hide thumbs Also See for Comet:
Table of Contents

Advertisement

Setting the potentiometer to the middle
position:
Loosen the bolts (1) on the drive lever
mounting.
Twist the potentiometer shaft (2) until the
same resistance (approx. 5kOhm) is
measured between the central connection
cable (3) and both outputs (4).
Retighten the bolts on the drive lever
mounting.
Check settings and repeat the adjustment if
necessary.
Reassembly of the operating console takes
place in reverse order
Check all functions (trial run).
45

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents