Not Charging - Sunrise Medical Rhapsody Service Manual

Table of Contents

Advertisement

1.4 Not Charging

If the chair is not charging through the rear charging
port,check that the circuit breaker is not triped. Insure the
three pin charger is properly connected and in good condi-
tion. Measure the voltage across pin one and pin two (fig
1.4.1)
If the voltage meter reads approximately 24 volts, replace
the charger, if the meter reads less than 12 volts, check the
coresponding pins ont the controller (fig 1.4.2) if the meter
reads less than 12 volts, proceed to the next step.
If all of the measurements read full voltage, then replace the controller.
SUNRISE MEDICAL
Section 1
Troubleshooting: No Power (cont.)
Troubleshooting: No Power (cont.)
RHAPSODY SERVICE MANUAL
A1.4.1
2006
PAGE 2
A1.4.2