The lower section of the dialog box contains the following
information:
7. DATA LOGGING
default setting is OFF. To turn on this feature, select ON for
Latching Alarm.
Voice: This feature enables voice messaging while
operating the instrument. The default setting is ON. To turn
off this feature, select OFF for Voice.
Confidence Beep: The confidence beep is a visual flash,
accompanied by an audible beep every 20 seconds to
reassure user that the instrument is working. The default
setting is OFF. To turn on this feature, select ON for
Confidence Beep.
Display Last Calibration Date: This feature allows the
date for the last calibration to be displayed in the start-up
routine each time the instrument is turned on. The default
setting is OFF. To turn on this feature, select ON for Display
Last Calibration Date.
Display Calibration Due Date: This feature allows the
calibration due date to be displayed in the start-up routine
each time the instrument is turned on. The default setting is
OFF. To turn on this feature, select ON for Display
Calibration Due Date.
Calibration Frequency: Calibration frequency represents
the number of days between calibrations. If Display
Calibration Due Date is selected, Calibration Frequency
must be specified. The default period is 30 days. To change
this option, choose between 1 and 90 days for Calibration
Frequency. After this period is chosen, the instrument must
be calibrated to initialize the process.
Gas type
Imm. Lo (Immediate low alarm levels)
Imm. Hi (Immediate high alarm levels)