APPENDIX J - CALIBRATING WITH MULTIPLE CANISTERS
b.
c.
d.
e.
f.
MPRO-4ABCH (O2, LEL, CO and SO2). Follow the
procedure in Section 5.2 Calibration Procedure (when all
calibration gases are in a single canister) because all the
calibration gases are in one canister (GFV279).
MPRO-2AC (O2 & CO). Follow the procedure in
Section 5.2 Calibration Procedure (when all calibration
gases are in a single canister) because all the calibration
gases are in a single canister (GFV123).
MPRO-2EF (NH3 and CL2). Follow the procedure in
Section 5.3
Calibration
calibration gases are in a single canister). The first gas is
NH3 (GFV213 or GFV250) and the second gas is CL2
(GFV252 or GFV249).
MPRO-AN (O2 and LEL/%by volume). Follow the
procedure in Section 5.2 Calibration Procedure (when all
calibration gases are in a single canister) because all the
gases are in a single cylinder (GFV123).
MPRO-4ABCJ (O2, LEL, CO and PH3). Follow the
procedure in Section 5.3 Calibration Procedure (when
not all calibration gases are in a single canister). The first
calibration gas is GFV123; the second calibration gas is
GFV272 (5PPM PH3).
Procedure
100
(when
not
all