GE EPS 3501 XL Operating Instructions Manual page 12

Electrophoresis power supply
Hide thumbs Also See for EPS 3501 XL:
Table of Contents

Advertisement

1
Introduction
1.3
Instrument
Keyboard
Key
set
ru n
pause
continue
ST OP
12
Label
set enter
ent er
-
run
pause
continue
STOP
Electrophoresis Power Supply EPS 3501 XL Operating Instructions 28-9639-69 AA
Description
Press set enter to enter a value or choice, confirm this if
valid, and move programming to the next field. Valid
values are voltage 35–3 500 V, current 1–400 mA, power
1–200 W, time 0:01–500 h, volthours 1–500 000 Vh,
milliampere-hours 1–25 000 mAh.
In the RUN mode, pressing set enter shows the
programmed parameters for the actual run. In addition,
set enter allows you to make changes in the program
during a run after first pressing pause continue.
After a run, when in END mode, pressing set enter puts
the instrument into SET, its programming mode.
to change the parameter, value or other
Press
choice in the field which is blinking. Numerical values are
changed in an accelerating manner when a key is held
down. Clicking a key changes the value in preset
increments. Parameters or units (e.g. Vh) and choices (e.g.
Yes/No) are changed with one key push. The keys can also
be used to switch between time and volthours in RUN,
PAUSE and END. The values scroll i.e. they automatically
change from maximum to minimum value or vice versa.
Press run to start the run and put the program into RUN
mode. The current values for voltage, current and power
are shown on the display. The elapsed time, volthours or
milliamperehours are also displayed. Switch between
these last three parameters with
Press pause continue to put the instrument in PAUSE
mode and switch off the voltage. The display shows the
status of the run at the time the key was pressed. pause
continue only operates in RUN mode. Time, integrated
voltage and integrated current are retained.
In the PAUSE MODE mode, set enter can be used to make
changes in the program. Return to RUN mode by pressing
pause continue or by pressing run.
Press STOP to stop the run and put the instrument in END
mode. The voltage is switched off and the end parameters
are displayed. Switch between time, integrated voltage
and integrated current by pressing
continued after pressing STOP. Press run to run the same
method again or press set enter to choose another
program, program a new method or make changes in an
existing method.
. A run cannot be