Section 8-Maintenance; Maintenance Safety Inspection Checklist; Hydraulic Pump - Invacare GET-U-UP LIFT Maintenance Manual

Invacare wheelchair/scooter lift owner's operator and maintenance manual
Table of Contents

Advertisement

SECTION 8—MAINTENANCE
•   When used for in‐home use, a full maintenance safety inspection is required prior to each new user.
•   If using the lift in the area of a shower or bath, ensure the lift is wiped clean of any moisture after use. Do not store the lift in a 
damp area or in a damp condition. Periodically inspect all components for signs of corrosion and replace those that are 
corroded or damaged.
•   The lift is designed to provide a maximum of safe, efficient and satisfactory service with minimum care and maintenance. If 
you question the safety of any part of the lift, contact your dealer immediately and advise them of the problem.
•   There is no adjustment or maintenance of the casters, other than cleaning, lubrication and checking axle and swivel bolts for 
tightness. Remove all debris, etc. from the caster and swivel bearings. If any parts are worn, replace these parts IMMEDIATELY.
•   Regular cleaning will reveal loose or worn parts, enhance smooth operation and extend the life expectancy of the lift.

Maintenance Safety Inspection Checklist

ITEM
THE CASTER BASE
Inspect for missing hardware.
Base opens/closes with ease.
Inspect casters and axle bolts for tightness.
Inspect casters for smooth swivel and roll.
Inspect and clear casters of debris.
Inspect roll pin to ensure secure to base.
THE MAST
Mast MUST be securely assembled to lift arms.
Inspect for bends or deflections.
Inspect pivot joints for wear.
THE LIFT ARMS AND LINKAGE
Check all hardware and attachment points.
Inspect for bends or deflections.
Inspect bolted joints of lift arms for wear.
Inspect to ensure that the lift arms are centered
between the base legs.
Ensure that the bolt is tightly secured.
Inspect pivot joints for wear.
SHIFTER HANDLE
Operates smoothly.
Locks adjustable base whenever engaged.
Check roll pin for wear
CLEANING
Whenever necessary.
SLINGS AND HARDWARE
Check all sling attachments for wear.
Inspect sling material for wear.
Inspect straps for wear.
Check that all labels are present and legible.

Hydraulic Pump

All parts of the hydraulic pump are precision machined then carefully assembled and tested to ensure reliable service. The pump 
assembly is completely enclosed and sealed with Neoprene rings to prevent leakage of hydraulic oil. A small amount of oil (about 
a drop) will accumulate around the piston from time to time and should be removed with a facial tissue.
The hydraulic pump is sealed at the factory and if service is required, the pump unit MUST be
returned to the factory for repair. DO NOT attempt to open the hydraulic pump or obtain local
service. This will void the warranty and might result in damage and costly repair. Consult your
dealer or contact Invacare for information.
Part No 1148115
INITIALLY
INSTITUTIONAL
INSPECT/ADJUST MONTHLY
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
WARNING
25
SECTION 8—MAINTENANCE
INSPECT/ADJUST PERIODICALLY
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
IN-HOME
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Get-U-Up
Lift

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ghs350Get-u-up

Table of Contents