Invacare Platinum IRC5LXO2 Operator's Manual page 32

Platinum series xl, 5, 10 oxygen concentrators standard, with senso2 and homefill ii
Hide thumbs Also See for Platinum IRC5LXO2:
Table of Contents

Advertisement

Available languages

Available languages

SECTION 7 —TROUBLESHOOTING GUIDE
SYMPTOM:
Alarm:
RAPID
Beep....Beep...
Beep...Beep
Platinum™Series
PROBABLE CAUSE:
LOW FLOW ALARM
1a. Kinked or blocked
tubing, cannula or
humidifier.
1b. Flowmeter set at 0.5
L/min. for Platinum 5 and
1 L/min. for Platinum 10
or less.
1a. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back On.
1b. Reset flowmeter to
prescribed flowrate. If
prescribed flowrate is below
0.5 L/min., use the Pediatric
Flowmeter IRCPF16 (Only
with Platinum 5 and XL).
32
SOLUTION:
Part No 1145728