SECTION 6 —MAINTENANCE
DO NOT operate the concentrator without the filters
installed.
FIGURE 6.1
Cabinet
Unplug the concentrator when cleaning. To avoid
electrical shock, DO NOT remove cabinet.
1. Clean the cabinet with a mild household cleaner and
non-abrasive cloth or sponge.
Humidifier
NOTE: To clean the oxygen humidifier, follow the instructions
provided by the manufacturer. If none are provided, follow these
steps:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten
(10) parts water and one (1) part vinegar.
3. Rinse thoroughly with hot water and refill with distilled
water to the level shown on the humidifier.
Platinum™Series
CAUTION
Cleaning the Cabinet Filter
WARNING
28
Cabinet Filter
(located on both
sides of
concentrator)
Part No 1145728