Step 3 - Shroud
Disassembly
a. Lift the center shroud from the main frame (figure A4.7).
Note: Center shroud is attached using a hook and loop fas-
tener system.
b. Use 4mm allen wrench to remove the fenders (figure
A4.8).
Reassembly
a. perform the reverse of instructions above.
Step 4 - Caster Arm
Disassembly
a. Use 6mm allen wrench and 17mm deep socket wrench to
remove the mounting bolt (fi gure A4.9).
Reassembly
a. perform the reverse of instructions above.
Adjustment
Note: The front caster arm can be adjusted from pivoting
slowly (fi gure A4.10).
a. Use a 4mm Allen wrench to back off the set screw and
use a 8mm bolt head (13mm Allen wrench) to loosen or
tighten the socket head bolt and make the proper adjust-
ment (operator preference).
SUNRISE MEDICAL
Disassembly/Reassembly, and Adjustment (cont)
QUICKIE S-11 SERVICE MANUAL
Socket Head Bolt
2005
A4.7
A4.8
A4.9
Set Screw
A4.10