Understanding VSI Controller Diagnostics Codes (cont.)
3.4 Four Bars- Right Motor Disconnected
The right hand motor has a bad connection. Check the connections to the right hand module.
Test Right Motor Open
Check that the batteries are fully charged and in good
condition, and check all cables and connections. Check
the connections to the right motor, look for a loose or dam-
aged connector. Remove the 9-pin Beau plug and check
the resistance across pin 1 and pin 2 as shown on (figure
A3.4.1). If the meter reads between 0 to 1.5 ohms, then
replace the controller.
If the meter reads between 0 and 1.5 ohms, then replace the controller.
Otherwise, check both brushes on the right motor. (Refer to
Section 4 Step 9 of this Service Manual for easiest access).
Ensure that the brushes are not excessively worn, (replace
as required). Top access for Motor brush shown in (figure
A3.4.2) (requires shroud removal).
Use the meter to check the resistance across the two
bottom contacts of the thicker wires on the 4-pin motor
connector as shown in (figure A3.4.3). If the meter reads
between 0 to 1.5 ohms, then replace the main harness. If
none of the above corrects the problem, replace the right
motor.
If the meter reads between 0 and 1.5 ohms, then replace the main harness. If this does not correct the prob-
lem, then replace the right motor.
SUNRISE MEDICAL
QUICKIE S-11 SERVICE MANUAL
1
3
5
7
9
2
4
6
8
2005
OFF
750
V DC
1000
V AC
200
200
20
200u
2000m
2000u
200m
20m
200m
20M
2000K
10A
200K
A DC
20K 2000
200
OHM
V mA
COM
10A
DIGITAL MULTIMETER
A3.4.1
A3.4.2
OFF
V DC
1000
750
V AC
200
200
20
200u
2000m
2000u
20m
200m
20M
200m
2000K
10A
200K
A DC
20K 2000
200
OHM
V mA
COM
10A
A3.4.3
DIGITAL MULTIMETER
PAGE 7

Advertisement

Table of Contents
loading

Table of Contents