Sunrise Medical Quickie S-11 Service Manual

Sunrise Medical Quickie S-11 Service Manual

Quickie s-11
Hide thumbs Also See for Quickie S-11:

Advertisement

Quickie S-11 Service Manual
©2005 Sunrise Medical Inc.
100740 Rev A

Advertisement

Table of Contents
loading

Summary of Contents for Sunrise Medical Quickie S-11

  • Page 1 Quickie S-11 Service Manual ©2005 Sunrise Medical Inc. 100740 Rev A...
  • Page 2: Table Of Contents

    Quickie S-11 Troubleshooting Guide INTRODUCTION ............0.1 Understanding VSI Controller Diagnostics Codes ................Section 3 Specifications VSI Controller ............0.2 3.1 1 Bar (1 Red) Low Battery Voltage ..... 5 Plugs/Connectors ............. 0.3 3.2 2 Bars (2 Red) - Left Motor Disconnected ..5 Main Wiring Diagram/ Tool List .........
  • Page 3: Introduction

    Service Manual. This Service Manual has been compiled as a troubleshooting guide for the Quickie S-11. Photographs and content may differ from the actual products in some cases due to changes in specifications and other factors.
  • Page 4: Vsi Controller

    Controller. Horn Key- Activates a warning horn. Speed/ Profile Decrease. Used to decrease the Speed/ Profile setting. Speed/ Profile Increase. Used to Increase the Speed/ Profile setting. CNTRLLR BTNS SMASP0P01 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 0.2...
  • Page 5: Plugs/Connectors

    LH Motor + LH Motor - 4-pin connect 1 = Brake Solenoid RH Motor - RH Motor + 2 = Brake Solenoid 3 = Motor 4 = Motor 9-pin Connect Harness Socket Face SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 0.3...
  • Page 6: Main Wiring Diagram/Tool List

    3. 5mm Allen wrench 4. 6mm Allen wrench 5. 13mm Open-end wrench 6. 13mm Socket wrench 7. 17mm Deep Socket wrench 8. Phillips screwdriver #2 9. Flat blade screwdriver 10. Cutter for the zip-tie SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 0.4...
  • Page 7: Troubleshooting: No Power

    If the voltage meter reads below 12volts,charge the batteries. Note: To find a bad battery, use a battery load tester. A1.2.1 If the voltage meter reads below 12 volts, charge the batteries. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 1...
  • Page 8: Step 1.3 Recharging The Batteries

    1000 V AC 200u 2000m 2000u 200m 200m 2000K 200K A DC 20K 2000 V mA DIGITAL MULTIMETER A1.5.2 If both of the measurements read full voltage, then replace the controller. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 2...
  • Page 9: Step 1.6 Battery Wire Harness

    If the meter reads more than 1 ohm, then change the circuit breaker. 1.9 Main Harness If the above steps did not correct the problem, change the main harness. If the previous steps did not correct the problem, change the main harness. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 3...
  • Page 10: Maximum Speed Indicator Ripples

    The control system safety circuits have been activated and the control system hasbeen prevented from moving the wheelchair. This indicates a system trip, i.e. the VSI has detected a problem somewhere in the wheelchair's electrical system. Please refer to Section 3 (VSI Controller Diagnostics). SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 4...
  • Page 11: Understanding Vsi Controller Diagnostics Codes

    V mA DIGITAL MULTIMETER A3.2.3 If the meter reads between 0 to 1.5 ohms, then replace the main harness. If none of the above corrects the problem, replace the left motor. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 5...
  • Page 12: Bars (3 Red) - Left Motor Wiring Trip

    Test 1 Test 2 Test 3 Test 4 If all of the readings are open, then replace the main harness. If any of the readings are short, then replace the left motor. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 6...
  • Page 13: Right Motor Disconnected

    V mA A3.4.3 DIGITAL MULTIMETER If the meter reads between 0 and 1.5 ohms, then replace the main harness. If this does not correct the prob- lem, then replace the right motor. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 7...
  • Page 14: Right Motor Wiring Trip

    Test 1 Test 2 Test 3 Test 4 If all of the readings are open, then replace the main harness. If any of the readings are short, then replace the right motor. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 8...
  • Page 15: Possible Joystick Trip

    V DC 1000 V AC 200u 2000m 2000u 200m 200m 2000K 200K A DC 20K 2000 V mA DIGITAL MULTIMETER A3.8.2 If both readings are approximately 30 ohms then replace the controller SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 9...
  • Page 16: High Battery Voltage

    A3.9.1 check the charger. Otherwise, replace your controller. If the meter reads more than 30 volts replace the charger, If the Batteries, connections, and voltage level are correct replace the controller. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 10...
  • Page 17: Disassembly/Reassembly

    Make sure the latching levers are locked flat against the seat post brackets. c. Plug the controller plugs into the plug receptacles at the rear of the chair. d. Install the Front Seat Post Knobs (figure A4.3). A4.3 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 11...
  • Page 18: Step 2 (Seat Posts)

    Pull out the three remaining quick release pins, remove all other seat posts (figure A4.6). A4.5 Reassembly a. perform the reverse of instructions above. A4.6 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 12...
  • Page 19: Step 3 (Shroud)

    Use a 4mm Allen wrench to back off the set screw and Set Screw use a 8mm bolt head (13mm Allen wrench) to loosen or tighten the socket head bolt and make the proper adjust- ment (operator preference). Socket Head Bolt A4.10 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 13...
  • Page 20: Step 5 (Charger)

    Slide the female charger plug out of the beau plug bracket (figure A4.12) c. Depress the locking tabs and unplug the A/C plug and charger light indicator plug. Reassembly a. Perform the reverse of instructions above. A4.11 A4.12 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 14...
  • Page 21: Step 6 (Battery)

    Depress the locking tabs and unplug the battery connec- tors. e. Grip the battery handle and pull the battery out base. Reassembly a. Perform the reverse of instructions above. A4.16 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 15...
  • Page 22: Step 7 (Motor/Wheel)

    Note: There is a suspension bumper on the frame that lim- the top frame’s movement (figure A4.20). a. Use a 13mm open wrench to loosen or tighten the jam nut and make the proper adjustment (operator preference). A4.20 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 16...
  • Page 23: Step 8 (Check/Replace Motor Brushes)

    A4.21 Hint: Make sure that the brushes are seated properly before re-installing the brush caps. Reassembly a. Reassemble your S11, by performing the Reverse of steps 6b. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 17...
  • Page 24: Step 9 (Wire Harnesses)

    Note: The circuit breaker can be removed without removing the main harness. a. Pull the quick disconnect terminal out from the bottom of the circuit breaker, then remove the circuit breaker locking A4.23 nut. A4.24 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 18...