Service Manual. This Service Manual has been compiled as a troubleshooting guide for the Quickie S-11. Photographs and content may differ from the actual products in some cases due to changes in specifications and other factors.
Controller. Horn Key- Activates a warning horn. Speed/ Profile Decrease. Used to decrease the Speed/ Profile setting. Speed/ Profile Increase. Used to Increase the Speed/ Profile setting. CNTRLLR BTNS SMASP0P01 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 0.2...
LH Motor + LH Motor - 4-pin connect 1 = Brake Solenoid RH Motor - RH Motor + 2 = Brake Solenoid 3 = Motor 4 = Motor 9-pin Connect Harness Socket Face SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 0.3...
If the voltage meter reads below 12volts,charge the batteries. Note: To find a bad battery, use a battery load tester. A1.2.1 If the voltage meter reads below 12 volts, charge the batteries. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 1...
1000 V AC 200u 2000m 2000u 200m 200m 2000K 200K A DC 20K 2000 V mA DIGITAL MULTIMETER A1.5.2 If both of the measurements read full voltage, then replace the controller. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 2...
If the meter reads more than 1 ohm, then change the circuit breaker. 1.9 Main Harness If the above steps did not correct the problem, change the main harness. If the previous steps did not correct the problem, change the main harness. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 3...
The control system safety circuits have been activated and the control system hasbeen prevented from moving the wheelchair. This indicates a system trip, i.e. the VSI has detected a problem somewhere in the wheelchair's electrical system. Please refer to Section 3 (VSI Controller Diagnostics). SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 4...
V mA DIGITAL MULTIMETER A3.2.3 If the meter reads between 0 to 1.5 ohms, then replace the main harness. If none of the above corrects the problem, replace the left motor. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 5...
Test 1 Test 2 Test 3 Test 4 If all of the readings are open, then replace the main harness. If any of the readings are short, then replace the left motor. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 6...
V mA A3.4.3 DIGITAL MULTIMETER If the meter reads between 0 and 1.5 ohms, then replace the main harness. If this does not correct the prob- lem, then replace the right motor. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 7...
Test 1 Test 2 Test 3 Test 4 If all of the readings are open, then replace the main harness. If any of the readings are short, then replace the right motor. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 8...
V DC 1000 V AC 200u 2000m 2000u 200m 200m 2000K 200K A DC 20K 2000 V mA DIGITAL MULTIMETER A3.8.2 If both readings are approximately 30 ohms then replace the controller SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 9...
A3.9.1 check the charger. Otherwise, replace your controller. If the meter reads more than 30 volts replace the charger, If the Batteries, connections, and voltage level are correct replace the controller. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 10...
Make sure the latching levers are locked flat against the seat post brackets. c. Plug the controller plugs into the plug receptacles at the rear of the chair. d. Install the Front Seat Post Knobs (figure A4.3). A4.3 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 11...
Pull out the three remaining quick release pins, remove all other seat posts (figure A4.6). A4.5 Reassembly a. perform the reverse of instructions above. A4.6 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 12...
Use a 4mm Allen wrench to back off the set screw and Set Screw use a 8mm bolt head (13mm Allen wrench) to loosen or tighten the socket head bolt and make the proper adjust- ment (operator preference). Socket Head Bolt A4.10 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 13...
Slide the female charger plug out of the beau plug bracket (figure A4.12) c. Depress the locking tabs and unplug the A/C plug and charger light indicator plug. Reassembly a. Perform the reverse of instructions above. A4.11 A4.12 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 14...
Depress the locking tabs and unplug the battery connec- tors. e. Grip the battery handle and pull the battery out base. Reassembly a. Perform the reverse of instructions above. A4.16 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 15...
Note: There is a suspension bumper on the frame that lim- the top frame’s movement (figure A4.20). a. Use a 13mm open wrench to loosen or tighten the jam nut and make the proper adjustment (operator preference). A4.20 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 16...
A4.21 Hint: Make sure that the brushes are seated properly before re-installing the brush caps. Reassembly a. Reassemble your S11, by performing the Reverse of steps 6b. SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 17...
Note: The circuit breaker can be removed without removing the main harness. a. Pull the quick disconnect terminal out from the bottom of the circuit breaker, then remove the circuit breaker locking A4.23 nut. A4.24 SUNRISE MEDICAL QUICKIE S-11 SERVICE MANUAL 2005 PAGE 18...